Decision Summary
Upload-ready calibration protocol overview
Instrument
P205s1
Session
2
Location
GreenbeltMD
Calibration person
Bryan Place
CF file
Pandora205s1_CF_v2d20231113.txt
Software version
p1-8
Calibration vector
-
Reference day
20231119
Overall status
development
O3Ext: YesO3Synt: YesNO2: YesSO2: With CareHCHO: YesH2O: Not answered
Action Timeline
99 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 27 | 2 | 0 | 10m 35s |
| AnalyzeDark | 10 | 1 | 0 | 4m 23s |
| AssignCalFile | 4 | 1 | 0 | 48s |
| CheckFilters | 4 | 1 | 0 | 1m 10s |
| CheckNoiseWave | 7 | 1 | 0 | 4m 25s |
| Convolution | 16 | 5 | 0 | 1m 7s |
| CreateReference | 74 | 37 | 14 | 14m 6s |
| FieldCalibration | 162 | 38 | 2 | 1h 56m |
| NDFilters | 16 | 1 | 0 | 1m 56s |
| No action | 17 | 0 | 0 | 0s |
| SensitivityCalibration | 64 | 3 | 0 | 4m 8s |
| StrayLight | 22 | 3 | 1 | 1m 28s |
| TempDependence | 7 | 1 | 0 | 1m 13s |
| WavelengthCalibration | 19 | 5 | 0 | 1m 57s |
Session Figures
11 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora205s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 116 KBCreate reference SELSPECS
Pandora205s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 81 KBCreate reference SLERR (STRAY)
Pandora205s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 82 KBCreate reference AODEST
Pandora205s1_CreateReference_AODEST.jpeg - 38 KBCreate reference SENSCORRFIN (SENS)
Pandora205s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 84 KBCreate reference DISPFIT (DISP)
Pandora205s1_CreateReference_DISPFIT(DISP).jpeg - 76 KBCreate reference WLCHANGE (DISP)
Pandora205s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 130 KBCreate reference SHIFT (CLOS)
Pandora205s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 76 KB
field calibration
O3
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
Both ous and out2 retrievals looked good. Ous was offset from ref instrument due to slant column offset, wrms for out is below limits
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
Yes
Calibrated with Ref method. Retrievals looks good, a little noise in wrms
SO2
Decision for SO2
With Care
Calibrated with ref method and columns had some inverse-u shape
HCHO
Decision for HCHO
Yes
Calibrated with Ref method and retrievals matched ref instrument well
H2O
Decision for H2O