Decision Summary
Upload-ready calibration protocol overview
Instrument
P204s1
Session
81
Location
-
Calibration person
-
CF file
-
Software version
1.8.58
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: UnusableO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Yes
Action Timeline
No start/finish intervals found.No timeline intervals found
Duration by Step
No completed durations found
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| No actions loaded | ||||
Session Figures
No session figures found for the loaded instrument/session.No figures loaded
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Unusable
O3T is at 241.5(climatology: 231!), initial O3Synt outside range
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
EMLE looks good
SO2
Decision for SO2
Yes
as usual: possibly slightly overestimated refSC required for avoiding negative SCsas usual: possibly slightly overestimated refSC required for avoiding negative SCs -looks ok in summer - too low in Winter.
HCHO
Decision for HCHO
Yes
looks good into 2024 but with time approaches sky data.
H2O
Decision for H2O
Yes
Sessions 8 80 and 81 were carried out for improving quality of wvt1. Seems to be ok.