Decision Summary

Upload-ready calibration protocol overview
Instrument P204s1
Session 81
Location -
Calibration person -
CF file -
Software version 1.8.58
Calibration vector -
Reference day -
Overall status development
O3Ext: UnusableO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Yes
Instrument P204s1
Session 81
Events 0
Figures 0
Action time 0s

Action Timeline

No start/finish intervals found.
No timeline intervals found

Duration by Step

No completed durations found

Action Summary

Action Events Runs Aborted Time
No actions loaded

Session Figures

No session figures found for the loaded instrument/session.
No figures loaded

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable
O3T is at 241.5(climatology: 231!), initial O3Synt outside range

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
EMLE looks good

SO2

Decision for SO2
Yes
as usual: possibly slightly overestimated refSC required for avoiding negative SCsas usual: possibly slightly overestimated refSC required for avoiding negative SCs -looks ok in summer - too low in Winter.

HCHO

Decision for HCHO
Yes
looks good into 2024 but with time approaches sky data.

H2O

Decision for H2O
Yes
Sessions 8 80 and 81 were carried out for improving quality of wvt1. Seems to be ok.