Decision Summary

Upload-ready calibration protocol overview
Instrument P200s1
Session 3
Location GrandForksND
Calibration person Markus Kilian
CF file Pandora200s1_CF_v3d20231207.txt
Software version 1.8.65
Calibration vector -
Reference day 20240630
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: YesH2O: Yes
Instrument P200s1
Session 3
Events 258
Figures 17
Action time 1h 12m

Action Timeline

76 action intervals
Nov 14, 202401:14 PM03:59 PMWavelengthCalibrationCreateReferenceFieldCalibrationAssignCalFileWavelengthCalibration 8s finishedCreateReference 14m 8s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 17s finishedCreateReference 0s abortedCreateReference 54s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 21s abortedCreateReference 58s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 19s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 19s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 18s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 11s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 21s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 42s finishedCreateReference 0s finishedFieldCalibration 1m 45s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 1m 21s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 19m 41s abortedFieldCalibration 20m 29s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 4m 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 4m 42s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 1m 7s finished

Duration by Step

FieldCalibration51m 58sCreateReference18m 48sAssignCalFile1m 7sWavelengthCalibration8s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 1m 7s
CreateReference 70 34 10 18m 48s
FieldCalibration 172 40 1 51m 58s
No action 10 0 0 0s
WavelengthCalibration 3 1 0 8s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora200s1_WavelengthCalibration_SlitFuncParam.jpeg - 125 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora200s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 115 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora200s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 86 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora200s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 87 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora200s1_CreateReference_AODEST.jpeg - 46 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora200s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 80 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora200s1_CreateReference_DISPFIT(DISP).jpeg - 87 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora200s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 89 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora200s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 113 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora200s1_FieldCalibration_VC_O3.jpeg - 79 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora200s1_FieldCalibration_SCREFINT_O3.jpeg - 52 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.917e-03; wrmsmax=2.334e+00; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=80; timeaver(min)=0.00; redu=2; scref=(1.063e-04,8.928e-05,9.084e-05); uscref=(9.276e-06,2.017e-06,8.911e-06)

FieldCalibration MLE NO2

Pandora200s1_FieldCalibration_MLE_NO2.jpeg - 121 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora200s1_FieldCalibration_VCMLE_NO2.jpeg - 81 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.396e-02; wrmsmax=7.414e-02; atvarmax(%)=4.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(7.537e-05); uscref=(5.159e-05)

FieldCalibration MLE SO2

Pandora200s1_FieldCalibration_MLE_SO2.jpeg - 60 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora200s1_FieldCalibration_VCMLE_SO2.jpeg - 63 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=8.711e-04; wrmsmax=6.425e-03; atvarmax(%)=3.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(1.676e-04); uscref=(2.038e-05)

FieldCalibration MLE HCHO

Pandora200s1_FieldCalibration_MLE_HCHO.jpeg - 64 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora200s1_FieldCalibration_VCMLE_HCHO.jpeg - 85 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
ok

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
ok

SO2

Decision for SO2
Yes
ok

HCHO

Decision for HCHO
Yes
ok

H2O

Decision for H2O
Yes
ok