Decision Summary

Upload-ready calibration protocol overview
Instrument P199s1
Session 3
Location -
Calibration person CWaldauf
CF file -
Software version 1.8.56
Calibration vector -
Reference day -
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P199s1
Session 3
Events 0
Figures 0
Action time 0s

Action Timeline

No start/finish intervals found.
No timeline intervals found

Duration by Step

No completed durations found

Action Summary

Action Events Runs Aborted Time
No actions loaded

Session Figures

No session figures found for the loaded instrument/session.
No figures loaded

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
P199 vs reference (P2) slope of 0.976 and daily columns matched excellently with one another

O3Synt

Decision for O3Synt

NO2

Decision for NO2
P199 vs reference (P2) slope of 1.06 and daily columns matched excellently with one another

SO2

Decision for SO2
P199 vs reference (P2) slope forced through 1 to get columns to match OK with one another

HCHO

Decision for HCHO
P199 vs reference (P2) slope of 1.13 and daily columns matched OK with one another

H2O

Decision for H2O