Decision Summary

Upload-ready calibration protocol overview
Instrument P198s1
Session 5
Location -
Calibration person MGebetsberger
CF file -
Software version 1.8.60
Calibration vector -
Reference day -
Overall status development
O3Ext: YesO3Synt: With CareNO2: With CareSO2: With CareHCHO: With CareH2O: Not answered
Instrument P198s1
Session 5
Events 0
Figures 0
Action time 0s

Action Timeline

No start/finish intervals found.
No timeline intervals found

Duration by Step

No completed durations found

Action Summary

Action Events Runs Aborted Time
No actions loaded

Session Figures

No session figures found for the loaded instrument/session.
No figures loaded

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
O3Synt shows some w shapes throughout day and AXC-T was estimated as >10K above climatology. O3ext columns and rms look good

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
With Care
Calibrated using Ref method. Followed Ref NO2 columns well, but NO2 rms a little high and noisy

SO2

Decision for SO2
With Care
Calibrated using Ref method. Followed Ref SO2 columns well, but rms was high

HCHO

Decision for HCHO
With Care
Calibrated using Ref method. Followed Ref HCHO columns OK, but rms was high

H2O

Decision for H2O