Decision Summary
Upload-ready calibration protocol overview
Instrument
P198s1
Session
4
Location
-
Calibration person
MGebetsberger
CF file
-
Software version
1.8.29
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: YesO3Synt: With CareNO2: UnusableSO2: YesHCHO: YesH2O: Yes
Action Timeline
No start/finish intervals found.No timeline intervals found
Duration by Step
No completed durations found
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| No actions loaded | ||||
Session Figures
No session figures found for the loaded instrument/session.No figures loaded
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
no extrapolation, good temperature uncertainty using cto0 temperature calibration code
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Unusable
up to 2DU, no background day availalbe, therefore rather high common uncertainty which could not be reduced with the E-MLE
SO2
Decision for SO2
Yes
close to zero with minor events
HCHO
Decision for HCHO
Yes
short period but systematic higher total columns compared to trop
H2O
Decision for H2O
Yes
Similar to P128 beiing one of the first units with H2O calibrated; bias to tropospheric H2O