Decision Summary
Upload-ready calibration protocol overview
Instrument
P196s1
Session
8
Location
GreenbeltMD
Calibration person
Bryan Place
CF file
Pandora196s1_CF_v8d20230627.txt
Software version
1.8.57
Calibration vector
-
Reference day
20230703
Overall status
development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Action Timeline
61 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 27s |
| CreateReference | 51 | 26 | 9 | 15m 49s |
| FieldCalibration | 93 | 34 | 2 | 57m 39s |
| No action | 14 | 0 | 0 | 0s |
Session Figures
8 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference SELSPECS
Pandora196s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 73 KBCreate reference SLERR (STRAY)
Pandora196s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 66 KBCreate reference SENSCORRFIN (SENS)
Pandora196s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 70 KBCreate reference DISPFIT (DISP)
Pandora196s1_CreateReference_DISPFIT(DISP).jpeg - 57 KBCreate reference WLCHANGE (DISP)
Pandora196s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 112 KBCreate reference SHIFT (CLOS)
Pandora196s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 75 KB
field calibration
O3
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
P196 vs reference (P2) slope of 0.99 and daily columns matched very well with one another
O3Synt
Decision for O3Synt
NO2
Decision for NO2
P196 vs reference (P2) slope of 1.08 and daily columns matched very well with one another
SO2
Decision for SO2
P196 vs reference (P2) slope of 1.15 and few filtered points remaining matched very well with one another
HCHO
Decision for HCHO
P196 vs reference (P2) slope of 1.04 and few filtered points remaining matched very well with one another
H2O
Decision for H2O