Decision Summary

Upload-ready calibration protocol overview
Instrument P196s1
Session 1
Location LabSciGlob
Calibration person Manuel Gebetsberger
CF file -
Software version 1.8.49
Calibration vector -
Reference day -
Overall status development
O3Ext: UnusableO3Synt: With CareNO2: With CareSO2: YesHCHO: With CareH2O: Unusable
Instrument P196s1
Session 1
Events 11
Figures 1
Action time 1m 22s

Action Timeline

3 action intervals
Sep 28, 202202:22 PM02:40 PMAnalyzeDarkAnalyzeBrightAnalyzeDark 11s abortedAnalyzeBright 56s abortedAnalyzeDark 15s aborted

Duration by Step

AnalyzeBright56sAnalyzeDark26s

Action Summary

Action Events Runs Aborted Time
AnalyzeBright 3 1 1 56s
AnalyzeDark 4 2 2 26s
No action 4 0 0 0s

Session Figures

1 figure from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora196s1_WavelengthCalibration_SlitFuncParam(A2).jpeg - 72 KB
Slit function parameters

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable
O3 Ext flagged out entirely

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
With Care
Only valid around the reference day

SO2

Decision for SO2
Yes
Used redu = 0

HCHO

Decision for HCHO
With Care
Only valid around the reference day

H2O

Decision for H2O
Unusable
All flagged out