Decision Summary
Upload-ready calibration protocol overview
Instrument
P190s1
Session
8
Location
-
Calibration person
-
CF file
-
Software version
1.8.26
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: UnusableO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Not answered
Action Timeline
No start/finish intervals found.No timeline intervals found
Duration by Step
No completed durations found
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| No actions loaded | ||||
Session Figures
No session figures found for the loaded instrument/session.No figures loaded
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Unusable
out2 O3 systematic invers-U, maybe to SLC issue, but L4 looks very good. increasing O3T over the day, but with two humps symmetric around noon.
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
up to 1e-3 NO2, wrms strongly drifting away from the reference day which causes flagging to DQ low
SO2
Decision for SO2
Yes
no M shape; increasing values over the day, events up to 2e-4
HCHO
Decision for HCHO
Yes
follows nicely the trop data with a systematic higher value. trop data look very good
H2O
Decision for H2O