Decision Summary

Upload-ready calibration protocol overview
Instrument P185s1
Session 4
Location EastProvidenceRI
Calibration person Stefanie Morhenn
CF file Pandora185s1_CF_v4d20210421.txt
Software version 1.8.55
Calibration vector -
Reference day 20220917
Overall status development
O3Ext: With CareO3Synt: UnusableNO2: With CareSO2: YesHCHO: With CareH2O: Yes
Instrument P185s1
Session 4
Events 428
Figures 17
Action time 7h 1m

Action Timeline

96 action intervals
May 12, 2023May 15, 2023Jun 07, 2023Jun 12, 202310:26 AMCreateReferenceFieldCalibrationAssignCalFileCreateReference 27s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 39s finishedCreateReference 0s abortedCreateReference 1m 9s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 56s finishedCreateReference 0s abortedCreateReference 43s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 1s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 59s abortedCreateReference 1m 1s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 35s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 34s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 57s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 41s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 22s finishedCreateReference 0s finishedCreateReference 6s abortedCreateReference 5s abortedFieldCalibration 5m 47s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 2h 40m finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 50m 47s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 23m 15s finishedFieldCalibration 0s abortedFieldCalibration 4s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 18s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 21m 3s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 25s finishedFieldCalibration 0s finishedAssignCalFile 3s abortedAssignCalFile 56s finishedFieldCalibration 6s abortedFieldCalibration 6s abortedFieldCalibration 9s abortedFieldCalibration 2h 20m abortedFieldCalibration 9s abortedFieldCalibration 16s abortedFieldCalibration 1m 43s abortedFieldCalibration 2m 23s abortedFieldCalibration 13s abortedFieldCalibration 5s abortedFieldCalibration 2s aborted

Duration by Step

FieldCalibration6h 48mCreateReference12m 15sAssignCalFile1m 0s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 5 2 1 1m 0s
CreateReference 87 42 14 12m 15s
FieldCalibration 319 52 13 6h 48m
No action 17 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora185s1_WavelengthCalibration_SlitFuncParam.jpeg - 238 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora185s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 172 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora185s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 134 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora185s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 118 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora185s1_CreateReference_AODEST.jpeg - 57 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora185s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 122 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora185s1_CreateReference_DISPFIT(DISP).jpeg - 109 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora185s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 114 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora185s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 161 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora185s1_FieldCalibration_VC_O3.jpeg - 98 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora185s1_FieldCalibration_SCREFINT_O3.jpeg - 79 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.019e-03; wrmsmax=9.000e-04; atvarmax(%)=2.000e+01; amflim=(1.150,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=18; timeaver(min)=0.00; redu=2; scref=(1.264e-04,1.054e-04,1.086e-04); uscref=(6.949e-06,2.851e-06,7.560e-06)

FieldCalibration MLE NO2

Pandora185s1_FieldCalibration_MLE_NO2.jpeg - 180 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora185s1_FieldCalibration_VCMLE_NO2.jpeg - 127 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.333e-02; wrmsmax=9.000e-04; atvarmax(%)=2.000e+01; amflim=(1.000,2.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(5.178e-05); uscref=(2.610e-05)

FieldCalibration MLE SO2

Pandora185s1_FieldCalibration_MLE_SO2.jpeg - 126 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora185s1_FieldCalibration_VCMLE_SO2.jpeg - 118 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.223e-03; wrmsmax=2.482e-02; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=11; timeaver(min)=0.00; redu=2; scref=(2.046e-04); uscref=(3.040e-05)

FieldCalibration MLE HCHO

Pandora185s1_FieldCalibration_MLE_HCHO.jpeg - 116 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora185s1_FieldCalibration_VCMLE_HCHO.jpeg - 127 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
O3Ext almost entirely flagegd out, O3Synt looks good, O3Temp has some wiggles / double values / zigzag, which can also be seen in wrms. Interpolation used in AXC

O3Synt

Decision for O3Synt
Unusable

NO2

Decision for NO2
With Care
NO2 overall a little bit too high (lowering the ref value by 4e-5 would lead to zero AMF-dependency. Was not possible to reach this value in field cal.) MLE used

SO2

Decision for SO2
Yes
looks good

HCHO

Decision for HCHO
With Care
noisy (due to noise in wrms?), sometimes below zero, seems to be some hardware issue that also reflects in noise in O3Temp. E-MLE used.

H2O

Decision for H2O
Yes
looks good