Decision Summary

Upload-ready calibration protocol overview
Instrument P178s1
Session 3
Location LabGSFC
Calibration person Bryan Place
CF file Pandora178s1_CF_v3d20240513.txt
Software version 1.8.60
Calibration vector -
Reference day 20240823
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P178s1
Session 3
Events 327
Figures 13
Action time 2h 40m

Action Timeline

66 action intervals
Oct 01, 2024Oct 02, 2024Nov 13, 2024Dec 02, 202402:57 PMCheckFiltersCheckNoiseWaveAnalyzeDarkAnalyzeBrightWavelengthCalibrationTempDependenceNDFiltersSensitivityCalibrationConvolutionCreateReferenceFieldCalibrationAssignCalFileCheckFilters 6s finishedCheckNoiseWave 17s finishedAnalyzeDark 1m 27s finishedAnalyzeBright 3m 47s finishedWavelengthCalibration 1m 32s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedTempDependence 46s finishedNDFilters 1m 22s finishedSensitivityCalibration 5m 48s finishedSensitivityCalibration 0s finishedSensitivityCalibration 0s finishedConvolution 1m 11s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 45s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 21s finishedCreateReference 0s abortedCreateReference 1m 33s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 57s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 34s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 3m 18s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 50s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 14s finishedCreateReference 0s finishedFieldCalibration 7m 16s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 2h 2m finishedFieldCalibration 0s abortedFieldCalibration 1m 22s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 28s finishedFieldCalibration 28s abortedAssignCalFile 2s aborted

Duration by Step

FieldCalibration2h 11mCreateReference11m 33sSensitivityCalibration5m 48sAnalyzeBright3m 47sWavelengthCalibration1m 32sAnalyzeDark1m 27sNDFilters1m 22sConvolution1m 11sTempDependence46sAssignCalFile30sCheckNoiseWave17sCheckFilters6s

Action Summary

Action Events Runs Aborted Time
AnalyzeBright 14 1 0 3m 47s
AnalyzeDark 10 1 0 1m 27s
AssignCalFile 5 2 1 30s
CheckFilters 4 1 0 6s
CheckNoiseWave 6 1 0 17s
Convolution 16 5 0 1m 11s
CreateReference 52 26 7 11m 33s
FieldCalibration 99 19 2 2h 11m
NDFilters 16 1 0 1m 22s
No action 15 0 0 0s
SensitivityCalibration 64 3 0 5m 48s
TempDependence 7 1 0 46s
WavelengthCalibration 19 5 0 1m 32s

Session Figures

13 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora178s1_WavelengthCalibration_SlitFuncParam.jpeg - 124 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora178s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 108 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora178s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 84 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora178s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 91 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora178s1_CreateReference_AODEST.jpeg - 38 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora178s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 82 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora178s1_CreateReference_DISPFIT(DISP).jpeg - 81 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora178s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 88 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora178s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 86 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora178s1_FieldCalibration_VC_O3.jpeg - 73 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora178s1_FieldCalibration_SCREFINT_O3.jpeg - 52 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=6.557e-03; wrmsmax=3.030e+00; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=50; timeaver(min)=0.00; redu=2; scref=(1.764e-04,1.445e-04,1.560e-04); uscref=(7.564e-06,2.500e-06,8.457e-06)

FieldCalibration MLE NO2

Pandora178s1_FieldCalibration_MLE_NO2.jpeg - 117 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora178s1_FieldCalibration_VCMLE_NO2.jpeg - 90 KB
FieldCalibration VCMLE NO2

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O