Decision Summary
Upload-ready calibration protocol overview
Instrument
P174s1
Session
4
Location
LabSciGlob
Calibration person
Ragi Rajagopalan
CF file
Pandora174s1_CF_v4d20230117.txt
Software version
1.8.57
Calibration vector
-
Reference day
20230721
Overall status
development
O3Ext: UnusableO3Synt: YesNO2: With CareSO2: With CareHCHO: With CareH2O: Not answered
Action Timeline
111 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 29s |
| Convolution | 16 | 5 | 0 | 35s |
| CreateReference | 69 | 34 | 14 | 13m 11s |
| FieldCalibration | 343 | 59 | 8 | 7h 56m |
| NDFilters | 2 | 1 | 1 | 33s |
| No action | 16 | 0 | 0 | 0s |
| SensitivityCalibration | 84 | 8 | 3 | 2m 40s |
| WavelengthCalibration | 5 | 2 | 1 | 1m 2s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora174s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 205 KBCreate reference SELSPECS
Pandora174s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 183 KBCreate reference SLERR (STRAY)
Pandora174s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 175 KBCreate reference AODEST
Pandora174s1_CreateReference_AODEST.jpeg - 95 KBCreate reference SENSCORRFIN (SENS)
Pandora174s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 159 KBCreate reference DISPFIT (DISP)
Pandora174s1_CreateReference_DISPFIT(DISP).jpeg - 149 KBCreate reference WLCHANGE (DISP)
Pandora174s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 294 KBCreate reference SHIFT (CLOS)
Pandora174s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 240 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Unusable
Similar to previous period
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
With Care
Similar to previous period
SO2
Decision for SO2
With Care
Similar to previous period
HCHO
Decision for HCHO
With Care
Similar to previous period
H2O
Decision for H2O