Decision Summary

Upload-ready calibration protocol overview
Instrument P173s1
Session 19
Location AtlantaGA
Calibration person Hanna Rohringer
CF file Pandora173s1_CF_v19d20231101.txt
Software version 1.8.74
Calibration vector calV2
Reference day -
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Yes
Instrument P173s1
Session 19
Events 276
Figures 17
Action time 4h 29m

Action Timeline

50 action intervals
Jan 29, 2025Feb 03, 2025Feb 04, 2025Feb 11, 202511:33 AMFieldCalibrationAssignCalFileFieldCalibration 1m 33s abortedFieldCalibration 9m 9s abortedFieldCalibration 5m 41s abortedFieldCalibration 2h 55m abortedFieldCalibration 6s abortedFieldCalibration 4m 5s finishedFieldCalibration 58s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 37m 38s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 3m 47s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 11s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s staleFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 0s staleFieldCalibration 25s finishedFieldCalibration 0s abortedFieldCalibration 18m 52s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 9m 59s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 1m 15s finished

Duration by Step

FieldCalibration4h 27mAssignCalFile1m 15s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 1m 15s
FieldCalibration 258 47 9 4h 27m
No action 15 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora173s1_WavelengthCalibration_SlitFuncParam.jpeg - 135 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora173s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 134 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora173s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 103 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora173s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 122 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora173s1_CreateReference_AODEST.jpeg - 46 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora173s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 114 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora173s1_CreateReference_DISPFIT(DISP).jpeg - 111 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora173s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 207 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora173s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 170 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora173s1_FieldCalibration_VC_O3.jpeg - 149 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora173s1_FieldCalibration_SCREFINT_O3.jpeg - 115 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=9.200e-04; wrmsmax=9.702e-04; atvarmax(%)=2.000e+01; amflim=(1.050,15.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=50; timeaver(min)=0.00; redu=2; scref=(1.815e-04,1.764e-04,1.664e-04); uscref=(1.830e-05,5.912e-06,1.929e-05)

FieldCalibration MLE NO2

Pandora173s1_FieldCalibration_MLE_NO2.jpeg - 250 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora173s1_FieldCalibration_VCMLE_NO2.jpeg - 135 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.189e-03; wrmsmax=2.403e-01; atvarmax(%)=2.000e+01; amflim=(1.670,2.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=8; timeaver(min)=0.00; redu=0; scref=(2.757e-05); uscref=(2.215e-05)

FieldCalibration MLE SO2

Pandora173s1_FieldCalibration_MLE_SO2.jpeg - 79 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora173s1_FieldCalibration_VCMLE_SO2.jpeg - 71 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.004e-04; wrmsmax=2.316e-02; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=8; timeaver(min)=0.00; redu=2; scref=(1.778e-04); uscref=(3.894e-05)

FieldCalibration MLE HCHO

Pandora173s1_FieldCalibration_MLE_HCHO.jpeg - 83 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora173s1_FieldCalibration_VCMLE_HCHO.jpeg - 102 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
AXCT 0.4 degree higher then out went with cto temp

O3Synt

Decision for O3Synt

NO2

Decision for NO2
if i want to get same scref as s14 i would need to go with MLE however i did proceed with NO2

SO2

Decision for SO2
probably not calibratable

HCHO

Decision for HCHO

H2O

Decision for H2O
Yes
ok