Decision Summary

Upload-ready calibration protocol overview
Instrument P173s1
Session 18
Location AtlantaGA
Calibration person Hanna Rohringer
CF file Pandora173s1_CF_v18d20230601.txt
Software version 1.8.74
Calibration vector calV2
Reference day -
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: YesH2O: Yes
Instrument P173s1
Session 18
Events 193
Figures 17
Action time 1h 18m

Action Timeline

42 action intervals
Jan 22, 2025Feb 24, 202508:21 AMFieldCalibrationCreateReferenceAssignCalFileFieldCalibration 45s abortedFieldCalibration 7s abortedFieldCalibration 4m 2s abortedFieldCalibration 55s abortedFieldCalibration 4m 19s finishedFieldCalibration 0s abortedFieldCalibration 15s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedCreateReference 8s abortedFieldCalibration 36m 44s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 21m 45s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 7m 2s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedCreateReference 2s abortedAssignCalFile 1m 4s finishedFieldCalibration 1m 15s aborted

Duration by Step

FieldCalibration1h 17mAssignCalFile1m 4sCreateReference10s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 1m 4s
CreateReference 4 2 2 10s
FieldCalibration 169 39 6 1h 17m
No action 17 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora173s1_WavelengthCalibration_SlitFuncParam.jpeg - 135 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora173s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 116 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora173s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 101 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora173s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 98 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora173s1_CreateReference_AODEST.jpeg - 46 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora173s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 93 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora173s1_CreateReference_DISPFIT(DISP).jpeg - 95 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora173s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 102 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora173s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 113 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora173s1_FieldCalibration_VC_O3.jpeg - 101 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora173s1_FieldCalibration_SCREFINT_O3.jpeg - 79 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=7.450e-04; wrmsmax=2.588e+00; atvarmax(%)=2.000e+01; amflim=(1.000,10.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=80; timeaver(min)=0.00; redu=2; scref=(1.232e-04,1.060e-04,1.156e-04); uscref=(1.238e-05,3.484e-06,1.211e-05)

FieldCalibration MLE NO2

Pandora173s1_FieldCalibration_MLE_NO2.jpeg - 146 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora173s1_FieldCalibration_VCMLE_NO2.jpeg - 96 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.709e-02; wrmsmax=2.334e-01; atvarmax(%)=2.000e+01; amflim=(1.050,2.250); amflimvar=(0.000,1.000); perc=0.00; ngroup=8; timeaver(min)=0.00; redu=0; scref=(2.203e-05); uscref=(2.039e-05)

FieldCalibration MLE SO2

Pandora173s1_FieldCalibration_MLE_SO2.jpeg - 88 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora173s1_FieldCalibration_VCMLE_SO2.jpeg - 78 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=3.374e-04; wrmsmax=1.845e-02; atvarmax(%)=2.000e+01; amflim=(1.200,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(3.297e-04); uscref=(3.281e-05)

FieldCalibration MLE HCHO

Pandora173s1_FieldCalibration_MLE_HCHO.jpeg - 68 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora173s1_FieldCalibration_VCMLE_HCHO.jpeg - 77 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
ok

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
ok

SO2

Decision for SO2
Yes
probably not calibratable since out is already bad went with an amf range from 1.23 - 2.35 MLe done with 230 data points. Data looked belly-shaped but they did the same in the previous period from AP - so probably itis what it is.

HCHO

Decision for HCHO
Yes
ok

H2O

Decision for H2O
Yes
ok