Decision Summary
Upload-ready calibration protocol overview
Instrument
P173s1
Session
14
Location
AtlantaGA
Calibration person
Apoorva Pandey
CF file
Pandora173s1_CF_v16d20231101.txt
Software version
1.8.57
Calibration vector
calV1
Reference day
20231119
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: With CareHCHO: YesH2O: Not answered
Action Timeline
130 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 11 | 4 | 1 | 3m 6s |
| CreateReference | 61 | 31 | 11 | 10m 26s |
| FieldCalibration | 296 | 95 | 9 | 4h 12m |
| No action | 9 | 0 | 0 | 0s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora173s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 134 KBCreate reference SELSPECS
Pandora173s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 103 KBCreate reference SLERR (STRAY)
Pandora173s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 122 KBCreate reference AODEST
Pandora173s1_CreateReference_AODEST.jpeg - 46 KBCreate reference SENSCORRFIN (SENS)
Pandora173s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 114 KBCreate reference DISPFIT (DISP)
Pandora173s1_CreateReference_DISPFIT(DISP).jpeg - 111 KBCreate reference WLCHANGE (DISP)
Pandora173s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 207 KBCreate reference SHIFT (CLOS)
Pandora173s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 170 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
O3-ext flagged out, O3T from AXC >10 K higher than climatology
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
E-MLE looks good
SO2
Decision for SO2
With Care
inverse U-shaped
HCHO
Decision for HCHO
Yes
E-MLE looks okay
H2O
Decision for H2O