Decision Summary

Upload-ready calibration protocol overview
Instrument P173s1
Session 10
Location LabSciGlob
Calibration person Apoorva Pandey
CF file Pandora173s1_CF_v12d20230601.txt
Software version 1.8.57
Calibration vector calV1
Reference day 20230724
Overall status development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: With CareHCHO: YesH2O: Not answered
Instrument P173s1
Session 10
Events 474
Figures 17
Action time 3h 28m

Action Timeline

95 action intervals
Nov 28, 2023 - Nov 29, 2023Nov 29, 2023Dec 07, 2023Dec 10, 2023Aug 22, 202401:46 PMWavelengthCalibrationNDFiltersSensitivityCalibrationConvolutionCreateReferenceFieldCalibrationAssignCalFileWavelengthCalibration 24s finishedWavelengthCalibration 0s abortedWavelengthCalibration 1m 28s abortedWavelengthCalibration 1m 32s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedNDFilters 57s finishedSensitivityCalibration 36s finishedSensitivityCalibration 0s finishedSensitivityCalibration 0s finishedConvolution 16s abortedConvolution 1m 59s finishedConvolution 0s finishedConvolution 2m 37s finishedCreateReference 37s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 13s abortedCreateReference 17s finishedCreateReference 0s abortedCreateReference 1m 57s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 3m 31s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 12s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 17s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 20s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 50s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 46s finishedCreateReference 0s finishedCreateReference 4s abortedFieldCalibration 18s abortedFieldCalibration 1m 6s finishedFieldCalibration 0s abortedFieldCalibration 7m 37s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 1h 9m abortedFieldCalibration 1m 50s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 1m 42s finishedFieldCalibration 0s abortedFieldCalibration 19s abortedFieldCalibration 1h 7m finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 13m 43s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 1m 46s finishedFieldCalibration 9s abortedFieldCalibration 5m 15s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 2m 6s abortedFieldCalibration 14s abortedAssignCalFile 1m 3s finishedFieldCalibration 4m 43s finishedFieldCalibration 0s abortedAssignCalFile 48s finishedFieldCalibration 0s staleFieldCalibration 6m 51s finishedFieldCalibration 0s aborted

Duration by Step

FieldCalibration3h 2mCreateReference12m 4sConvolution4m 52sAssignCalFile3m 37sWavelengthCalibration3m 24sNDFilters57sSensitivityCalibration36s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 9 3 0 3m 37s
Convolution 13 4 1 4m 52s
CreateReference 64 32 10 12m 4s
FieldCalibration 235 43 10 3h 2m
NDFilters 50 1 0 57s
No action 15 0 0 0s
SensitivityCalibration 63 3 0 36s
WavelengthCalibration 25 8 2 3m 24s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora173s1_WavelengthCalibration_SlitFuncParam.jpeg - 135 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora173s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 116 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora173s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 101 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora173s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 98 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora173s1_CreateReference_AODEST.jpeg - 46 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora173s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 93 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora173s1_CreateReference_DISPFIT(DISP).jpeg - 95 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora173s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 102 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora173s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 113 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora173s1_FieldCalibration_VC_O3.jpeg - 78 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora173s1_FieldCalibration_SCREFINT_O3.jpeg - 62 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=8.827e-04; wrmsmax=3.816e-04; atvarmax(%)=3.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=25; timeaver(min)=0.00; redu=2; scref=(1.319e-04,1.033e-04,1.148e-04); uscref=(1.718e-05,5.950e-06,1.713e-05)

FieldCalibration MLE NO2

Pandora173s1_FieldCalibration_MLE_NO2.jpeg - 134 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora173s1_FieldCalibration_VCMLE_NO2.jpeg - 98 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(102); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.045e-02; wrmsmax=3.053e-01; atvarmax(%)=7.000e+01; amflim=(1.000,3.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=3; timeaver(min)=0.00; redu=0; scref=(4.585e-05); uscref=(5.905e-05)

FieldCalibration MLE SO2

Pandora173s1_FieldCalibration_MLE_SO2.jpeg - 83 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora173s1_FieldCalibration_VCMLE_SO2.jpeg - 76 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(102); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.006e-04; wrmsmax=1.694e-03; atvarmax(%)=1.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=70.00; ngroup=3; timeaver(min)=0.00; redu=2; scref=(2.904e-04); uscref=(6.171e-05)

FieldCalibration MLE HCHO

Pandora173s1_FieldCalibration_MLE_HCHO.jpeg - 85 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora173s1_FieldCalibration_VCMLE_HCHO.jpeg - 104 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
O3-ext flagged out, O3T from AXC >10 K higher than climatology

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
E-MLE looks good

SO2

Decision for SO2
With Care
high uncertainty

HCHO

Decision for HCHO
Yes
E-MLE looks okay, flagged out toward the end of the VP due to resolution change

H2O

Decision for H2O