Decision Summary
Upload-ready calibration protocol overview
Instrument
P169s1
Session
2
Location
Egbert
Calibration person
Ilias Bougoudis
CF file
Pandora169s1_CF_v5d20201119.txt
Software version
p1-8
Calibration vector
-
Reference day
20201202-20210408
Overall status
development
O3Ext: UnusableO3Synt: UnusableNO2: With CareSO2: UnusableHCHO: UnusableH2O: Unusable
Action Timeline
191 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 4 | 2 | 2 | 13s |
| AnalyzeDark | 7 | 3 | 3 | 34s |
| AssignCalFile | 6 | 2 | 0 | 1m 54s |
| CheckFilters | 4 | 2 | 2 | 23s |
| CheckNoiseWave | 2 | 1 | 1 | 14s |
| CreateReference | 197 | 94 | 31 | 38m 52s |
| FieldCalibration | 438 | 75 | 9 | 4h 25m |
| NDFilters | 6 | 3 | 3 | 11s |
| No action | 46 | 0 | 0 | 0s |
| SensitivityCalibration | 4 | 2 | 2 | 10s |
| TempDependence | 2 | 1 | 1 | 2s |
| WavelengthCalibration | 13 | 6 | 6 | 1m 48s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora169s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 163 KBCreate reference SELSPECS
Pandora169s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 117 KBCreate reference SLERR (STRAY)
Pandora169s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 127 KBCreate reference AODEST
Pandora169s1_CreateReference_AODEST.jpeg - 53 KBCreate reference SENSCORRFIN (SENS)
Pandora169s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 110 KBCreate reference DISPFIT (DISP)
Pandora169s1_CreateReference_DISPFIT(DISP).jpeg - 97 KBCreate reference WLCHANGE (DISP)
Pandora169s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 123 KBCreate reference SHIFT (CLOS)
Pandora169s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 115 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Unusable
.......
O3Synt
Decision for O3Synt
Unusable
NO2
Decision for NO2
With Care
trends in wrms, wl_shift
SO2
Decision for SO2
Unusable
most of it flagged out
HCHO
Decision for HCHO
Unusable
flagged out
H2O
Decision for H2O
Unusable
flagged out