Decision Summary
Upload-ready calibration protocol overview
Instrument
P169s1
Session
1
Location
Egbert
Calibration person
Ilias Bougoudis
CF file
C:/Blick/data/calresults/Pandora169s1/session1/Pandora169s1_CF_session1.txt
Software version
1.8.49
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: UnusableO3Synt: UnusableNO2: With CareSO2: With CareHCHO: UnusableH2O: With Care
Action Timeline
113 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 40 | 6 | 4 | 33m 41s |
| AnalyzeDark | 110 | 20 | 7 | 16m 15s |
| CheckFilters | 24 | 8 | 5 | 2m 59s |
| CheckNoiseWave | 24 | 6 | 4 | 2m 19s |
| Convolution | 32 | 10 | 0 | 1m 15s |
| NDFilters | 108 | 3 | 1 | 13m 0s |
| No action | 31 | 0 | 0 | 0s |
| SensitivityCalibration | 205 | 25 | 21 | 55m 53s |
| TempDependence | 14 | 2 | 0 | 1m 59s |
| WavelengthCalibration | 111 | 33 | 16 | 20m 38s |
Session Figures
1 figure from Google Drive/figuresAutomatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Unusable
used dqf2in = [103]
O3Synt
Decision for O3Synt
Unusable
NO2
Decision for NO2
With Care
Used redu = 1
SO2
Decision for SO2
With Care
Used redu = 0
HCHO
Decision for HCHO
Unusable
Used redu = 0
H2O
Decision for H2O
With Care
H2O available although O3 was flagged out