Decision Summary

Upload-ready calibration protocol overview
Instrument P167s1
Session 14
Location -
Calibration person -
CF file -
Software version 1.8.56
Calibration vector -
Reference day -
Overall status development
O3Ext: UnusableO3Synt: With CareNO2: YesSO2: UnusableHCHO: With CareH2O: Unusable
Instrument P167s1
Session 14
Events 0
Figures 0
Action time 0s

Action Timeline

No start/finish intervals found.
No timeline intervals found

Duration by Step

No completed durations found

Action Summary

Action Events Runs Aborted Time
No actions loaded

Session Figures

No session figures found for the loaded instrument/session.
No figures loaded

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable
winter reference, seasonal wrms change

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes

SO2

Decision for SO2
Unusable
I didn't calibrate SO2 here because the MLE didn't make any sense and looked so bad i decided it was "uncalibratable". I think this is because this session uses a winter period and the others don't

HCHO

Decision for HCHO
With Care
same as session 9

H2O

Decision for H2O
Unusable