Decision Summary

Upload-ready calibration protocol overview
Instrument P167s1
Session 10
Location KenoshaWI
Calibration person Leonie Haunold
CF file Pandora167s1_CF_v10d20240614.txt
Software version 1.8.58
Calibration vector -
Reference day 20240902
Overall status development
O3Ext: UnusableO3Synt: With CareNO2: YesSO2: With CareHCHO: With CareH2O: Unusable
Instrument P167s1
Session 10
Events 282
Figures 17
Action time 7h 12m

Action Timeline

65 action intervals
Nov 16, 2024Nov 17, 2024Jan 27, 202506:18 PMCreateReferenceFieldCalibrationAssignCalFileCreateReference 9m 4s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 22s finishedCreateReference 0s abortedCreateReference 9m 27s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 27s abortedCreateReference 22s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 4s finishedCreateReference 0s abortedCreateReference 47s abortedCreateReference 27s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 5m 3s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 57s finishedCreateReference 0s finishedFieldCalibration 6m 13s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s staleFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 6h 28m finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 8m 12s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 2s abortedAssignCalFile 1m 22s finishedFieldCalibration 8s aborted

Duration by Step

FieldCalibration6h 42mCreateReference28m 0sAssignCalFile1m 22s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 1m 22s
CreateReference 55 29 10 28m 0s
FieldCalibration 211 34 2 6h 42m
No action 13 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora167s1_WavelengthCalibration_SlitFuncParam.jpeg - 106 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora167s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 122 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora167s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 88 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora167s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 89 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora167s1_CreateReference_AODEST.jpeg - 40 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora167s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 79 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora167s1_CreateReference_DISPFIT(DISP).jpeg - 67 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora167s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 134 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora167s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 106 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora167s1_FieldCalibration_VC_O3.jpeg - 71 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora167s1_FieldCalibration_SCREFINT_O3.jpeg - 53 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 redu=1 settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=3.555e-04; wrmsmax=2.565e+00; atvarmax(%)=2.000e+01; amflim=(1.600,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=100; timeaver(min)=0.00; redu=1; scref=(1.063e-04,8.844e-05,8.930e-05); uscref=(8.385e-06,1.995e-06,9.088e-06)

FieldCalibration MLE NO2

Pandora167s1_FieldCalibration_MLE_NO2.jpeg - 142 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora167s1_FieldCalibration_VCMLE_NO2.jpeg - 86 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=9.568e-03; wrmsmax=4.744e-01; atvarmax(%)=2.000e+01; amflim=(1.000,2.200); amflimvar=(0.000,1.000); perc=5.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(2.710e-05); uscref=(6.253e-06)

FieldCalibration MLE SO2

Pandora167s1_FieldCalibration_MLE_SO2.jpeg - 69 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora167s1_FieldCalibration_VCMLE_SO2.jpeg - 69 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=5.353e-03; wrmsmax=8.177e-02; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(6.767e-05); uscref=(1.916e-05)

FieldCalibration MLE HCHO

Pandora167s1_FieldCalibration_MLE_HCHO.jpeg - 71 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora167s1_FieldCalibration_VCMLE_HCHO.jpeg - 80 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable
okay

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes

SO2

Decision for SO2
With Care

HCHO

Decision for HCHO
With Care
same as session 9

H2O

Decision for H2O
Unusable