Decision Summary

Upload-ready calibration protocol overview
Instrument P162s1
Session 20
Location -
Calibration person -
CF file -
Software version 1.8.73
Calibration vector -
Reference day -
Overall status development
O3Ext: YesO3Synt: YesNO2: With CareSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P162s1
Session 20
Events 0
Figures 11
Action time 0s

Action Timeline

No start/finish intervals found.
No timeline intervals found

Duration by Step

No completed durations found

Action Summary

Action Events Runs Aborted Time
No actions loaded

Session Figures

11 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora162s1_WavelengthCalibration_SlitFuncParam.jpeg - 122 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora162s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 107 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora162s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 84 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora162s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 90 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora162s1_CreateReference_AODEST.jpeg - 38 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora162s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 84 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora162s1_CreateReference_DISPFIT(DISP).jpeg - 84 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora162s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 84 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora162s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 90 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora162s1_FieldCalibration_VC_O3.jpeg - 73 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora162s1_FieldCalibration_SCREFINT_O3.jpeg - 54 KB
FieldCalibration SCREFINT O3

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
AXC estimated O3T matched climatology. Both Out2 and Ous1 columns looked very good and agreed well with Greenbelt instruments

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
With Care
Checked wrms structure and looked good, did not calibrate NO2 for clearance

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O