Decision Summary

Upload-ready calibration protocol overview
Instrument P158s1
Session 2
Location GreenbeltMD
Calibration person Bryan Place
CF file Pandora158s1_CF_v2d20230421.txt
Software version p1-8
Calibration vector -
Reference day 20230618
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P158s1
Session 2
Events 229
Figures 10
Action time 3h 24m

Action Timeline

87 action intervals
Jun 13, 2023Jun 19, 2023Jul 17, 2023Jul 20, 2023CreateReferenceFieldCalibrationAssignCalFileCreateReference 1m 30s finishedCreateReference 0s abortedCreateReference 41s finishedCreateReference 0s abortedCreateReference 2m 10s finishedCreateReference 0s abortedCreateReference 1m 7s finishedCreateReference 0s abortedCreateReference 47s finishedCreateReference 0s abortedCreateReference 44s finishedCreateReference 0s abortedCreateReference 43s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 26s finishedCreateReference 0s abortedCreateReference 1m 42s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 4m 56s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 30s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 27s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 53s abortedCreateReference 19s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 35s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 53s finishedCreateReference 0s finishedFieldCalibration 2h 19m finishedFieldCalibration 0s abortedFieldCalibration 15s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 35s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 26m 19s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedCreateReference 6s abortedFieldCalibration 37s abortedFieldCalibration 28s abortedFieldCalibration 1m 28s abortedFieldCalibration 12s abortedFieldCalibration 1m 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 6m 59s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 6m 47s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 30s finished

Duration by Step

FieldCalibration3h 4mCreateReference19m 31sAssignCalFile30s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 30s
CreateReference 89 44 16 19m 31s
FieldCalibration 123 42 7 3h 4m
No action 14 0 0 0s

Session Figures

10 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora158s1_WavelengthCalibration_SlitFuncParam.jpeg - 94 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora158s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 114 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora158s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 103 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora158s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 90 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora158s1_CreateReference_AODEST.jpeg - 38 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora158s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 80 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora158s1_CreateReference_DISPFIT(DISP).jpeg - 78 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora158s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 123 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora158s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 92 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora158s1_FieldCalibration_VC_O3.jpeg - 72 KB
FieldCalibration VC O3

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
P158 vs reference (P2) slope of 0.99 and daily columns matched excellently with one another

O3Synt

Decision for O3Synt

NO2

Decision for NO2
P158 vs reference (P2) slope of 1.11 and daily columns matched very well with one another

SO2

Decision for SO2
P158 vs reference (P2) slope of 1 - used fitmeth 7 which forces slope to 1, this gave best overlap of columns

HCHO

Decision for HCHO
P158 vs reference (P2) slope of 0.93 and daily columns matched very well with one another

H2O

Decision for H2O