Decision Summary

Upload-ready calibration protocol overview
Instrument P158s1
Session 16
Location AtlantaGA
Calibration person Markus Kilian
CF file Pandora158s1_CF_v16d20190910.txt
Software version 1.8.70
Calibration vector -
Reference day 20190912
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P158s1
Session 16
Events 95
Figures 17
Action time 29m 40s

Action Timeline

32 action intervals
Jan 15, 202507:35 AM09:14 AMCreateReferenceFieldCalibrationAssignCalFileCreateReference 33s finishedCreateReference 0s abortedCreateReference 11s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 13s finishedCreateReference 0s abortedCreateReference 3m 35s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 20s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 25s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 27s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 25s finishedCreateReference 0s finishedFieldCalibration 21m 4s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 27s finished

Duration by Step

FieldCalibration21m 4sCreateReference8m 9sAssignCalFile27s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 27s
CreateReference 45 24 7 8m 9s
FieldCalibration 35 7 0 21m 4s
No action 12 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora158s1_WavelengthCalibration_SlitFuncParam.jpeg - 123 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora158s1_CreateReference_L1CHARFILT_OPEN.jpeg - 94 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora158s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 84 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora158s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 88 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora158s1_CreateReference_AODEST.jpeg - 51 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora158s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 82 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora158s1_CreateReference_DISPFIT(DISP).jpeg - 79 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora158s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 89 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora158s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 124 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora158s1_FieldCalibration_VC_O3.jpeg - 83 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora158s1_FieldCalibration_SCREFINT_O3.jpeg - 55 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 redu=1 settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.655e-03; wrmsmax=3.088e-01; atvarmax(%)=3.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=18; timeaver(min)=0.00; redu=1; scref=(2.714e-04,2.286e-04,2.529e-04); uscref=(9.163e-06,8.666e-06,8.497e-06)

FieldCalibration MLE NO2

Pandora158s1_FieldCalibration_MLE_NO2.jpeg - 134 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora158s1_FieldCalibration_VCMLE_NO2.jpeg - 74 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora158s1_FieldCalibration_MLE_SO2.jpeg - 61 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora158s1_FieldCalibration_VCMLE_SO2.jpeg - 53 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora158s1_FieldCalibration_MLE_HCHO.jpeg - 62 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora158s1_FieldCalibration_VCMLE_HCHO.jpeg - 74 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O