Decision Summary
Upload-ready calibration protocol overview
Instrument
P155s1
Session
1
Location
LabGSFC
Calibration person
Apoorva Pandey
CF file
Pandora155s1_CF_v1d20231018.txt
Software version
p1-8
Calibration vector
-
Reference day
20231024
Overall status
development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: YesH2O: Not answered
Action Timeline
97 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 17 | 2 | 1 | 3m 1s |
| AnalyzeDark | 10 | 1 | 0 | 1m 50s |
| AssignCalFile | 4 | 1 | 0 | 36s |
| CheckFilters | 6 | 2 | 1 | 8s |
| CheckNoiseWave | 6 | 1 | 0 | 1m 11s |
| Convolution | 16 | 5 | 0 | 44s |
| CreateReference | 58 | 29 | 10 | 12m 31s |
| FieldCalibration | 157 | 37 | 4 | 6h 14m |
| NDFilters | 30 | 4 | 3 | 1m 41s |
| No action | 11 | 0 | 0 | 0s |
| SensitivityCalibration | 91 | 4 | 1 | 5m 39s |
| StrayLight | 15 | 5 | 1 | 21s |
| TempDependence | 7 | 1 | 0 | 50s |
| WavelengthCalibration | 19 | 5 | 0 | 2m 7s |
Session Figures
11 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora155s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 147 KBCreate reference SELSPECS
Pandora155s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 127 KBCreate reference SLERR (STRAY)
Pandora155s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 126 KBCreate reference AODEST
Pandora155s1_CreateReference_AODEST.jpeg - 53 KBCreate reference SENSCORRFIN (SENS)
Pandora155s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 116 KBCreate reference DISPFIT (DISP)
Pandora155s1_CreateReference_DISPFIT(DISP).jpeg - 102 KBCreate reference WLCHANGE (DISP)
Pandora155s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 201 KBCreate reference SHIFT (CLOS)
Pandora155s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 148 KB
field calibration
O3
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
AXC looks good
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
Yes
Ref method, looks good
SO2
Decision for SO2
Yes
Ref method, looks good
HCHO
Decision for HCHO
Yes
Ref method, looks good
H2O
Decision for H2O