Decision Summary
Upload-ready calibration protocol overview
Instrument
P153s1
Session
3
Location
LabSciGlob
Calibration person
Manuel Gebetsberger
CF file
Pandora153s1_CF_v1d20190122.txt
Software version
1.8.26
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: UnusableO3Synt: With CareNO2: YesSO2: YesHCHO: With CareH2O: Not answered
Action Timeline
28 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 2 | 1 | 1 | 7s |
| AssignCalFile | 3 | 1 | 0 | 7m 26s |
| CheckDispersion | 22 | 7 | 0 | 14m 36s |
| CheckFilters | 4 | 1 | 0 | 1m 13s |
| CreateReference | 114 | 10 | 2 | 8m 49s |
| NDFilters | 38 | 3 | 2 | 9m 13s |
| No action | 11 | 0 | 0 | 0s |
| SensitivityCalibration | 34 | 4 | 1 | 5m 4s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora153s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 132 KBCreate reference SELSPECS
Pandora153s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 86 KBCreate reference SLERR (STRAY)
Pandora153s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 82 KBCreate reference AODEST
Pandora153s1_CreateReference_AODEST.jpeg - 37 KBCreate reference SENSCORRFIN (SENS)
Pandora153s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 91 KBCreate reference DISPFIT (DISP)
Pandora153s1_CreateReference_DISPFIT(DISP).jpeg - 114 KBCreate reference WLCHANGE (DISP)
Pandora153s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 87 KBCreate reference SHIFT (CLOS)
Pandora153s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 87 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Unusable
AXC did not look good, extrapol. needed, out looked good; ous data looked also okay but has a very wiggly wrms which slightly affect the data; temp looks bad has a w-shape
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
could not process tropdata, so MLE is used
SO2
Decision for SO2
Yes
looks good
HCHO
Decision for HCHO
With Care
could not process tropdata, so MLE is used; the first days have negative values, then there are a few good days, but then everything is flagged because of drift in the wavelength shift
H2O
Decision for H2O