Decision Summary
Upload-ready calibration protocol overview
Instrument
P150s1
Session
6
Location
-
Calibration person
-
CF file
-
Software version
1.8.26
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: With CareSO2: With CareHCHO: UnusableH2O: Not answered
Action Timeline
No start/finish intervals found.No timeline intervals found
Duration by Step
No completed durations found
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| No actions loaded | ||||
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora150s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 118 KBCreate reference SELSPECS
Pandora150s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 101 KBCreate reference SLERR (STRAY)
Pandora150s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 95 KBCreate reference AODEST
Pandora150s1_CreateReference_AODEST.jpeg - 38 KBCreate reference SENSCORRFIN (SENS)
Pandora150s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 101 KBCreate reference DISPFIT (DISP)
Pandora150s1_CreateReference_DISPFIT(DISP).jpeg - 149 KBCreate reference WLCHANGE (DISP)
Pandora150s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 114 KBCreate reference SHIFT (CLOS)
Pandora150s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 139 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
no extrap uncert 1.35e-3
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
With Care
EMLE, ok for the few times it was tracking
SO2
Decision for SO2
With Care
ok for the few times it was tracking
HCHO
Decision for HCHO
Unusable
MLE, high uncert, had to use DQF 103
H2O
Decision for H2O