Decision Summary
Upload-ready calibration protocol overview
Instrument
P149s1
Session
2
Location
-
Calibration person
LHaunold
CF file
-
Software version
1.8.49
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: UnusableO3Synt: UnusableNO2: With CareSO2: With CareHCHO: UnusableH2O: Yes
Action Timeline
No start/finish intervals found.No timeline intervals found
Duration by Step
No completed durations found
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| No actions loaded | ||||
Session Figures
1 figure from Google Drive/figuresAutomatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Unusable
Combination of bad instruemnt and bad day selection
O3Synt
Decision for O3Synt
Unusable
NO2
Decision for NO2
With Care
Combination of bad instruemnt and bad day selection
SO2
Decision for SO2
With Care
Combination of bad instruemnt and bad day selection
HCHO
Decision for HCHO
Unusable
Combination of bad instruemnt and bad day selection
H2O
Decision for H2O
Yes
only reasonable product