Decision Summary

Upload-ready calibration protocol overview
Instrument P145s1
Session 13
Location Toronto-Scarborough
Calibration person Markus Kilian
CF file Pandora145s1_CF_v14d20250711.txt
Software version 1.8.94
Calibration vector calV3
Reference day 20250715
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P145s1
Session 13
Events 293
Figures 4
Action time 3h 39m

Action Timeline

65 action intervals
Jun 26, 2026Jun 29, 202609:33 AM11:48 AMFieldCalibrationCreateReferenceAssignCalFileFieldCalibration 26m 21s abortedFieldCalibration 0s staleCreateReference 27m 24s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 27s finishedCreateReference 0s abortedCreateReference 1m 19s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 5s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 26s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 6s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 5m 38s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 42s finishedCreateReference 0s finishedCreateReference 1s abortedFieldCalibration 2m 1s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 1h 32m finishedFieldCalibration 0s abortedFieldCalibration 14m 42s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedCreateReference 1s abortedFieldCalibration 9m 32s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 32m 38s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 1m 2s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 2s abortedAssignCalFile 35s finished

Duration by Step

FieldCalibration2h 59mCreateReference40m 11sAssignCalFile35s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 35s
CreateReference 57 28 9 40m 11s
FieldCalibration 223 35 4 2h 59m
No action 10 0 0 0s

Session Figures

4 figures from Google Drive/figures
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=8.799e-03; wrmsmax=9.000e-04; atvarmax(%)=2.000e+01; amflim=(1.000,2.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(1.021e-04); uscref=(4.026e-05)

FieldCalibration MLE SO2

Pandora145s1_FieldCalibration_MLE_SO2.jpeg - 62 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora145s1_FieldCalibration_VCMLE_SO2.jpeg - 54 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=3.548e-04; wrmsmax=9.000e-04; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(2.461e-04); uscref=(3.932e-05)

FieldCalibration MLE HCHO

Pandora145s1_FieldCalibration_MLE_HCHO.jpeg - 67 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora145s1_FieldCalibration_VCMLE_HCHO.jpeg - 92 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O