Decision Summary

Upload-ready calibration protocol overview
Instrument P142s1
Session 16
Location -
Calibration person Hanna Rohringer
CF file Pandora142s1_CF_v16d20231028.txt
Software version 1.8.89
Calibration vector calV2
Reference day -
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: With CareH2O: Yes
Instrument P142s1
Session 16
Events 23
Figures 17
Action time 7m 42s

Action Timeline

5 action intervals
Mar 23, 2026Mar 24, 202601:55 PM08:45 AMFieldCalibrationAssignCalFileFieldCalibration 6m 54s finishedFieldCalibration 0s abortedFieldCalibration 5s abortedAssignCalFile 43s finishedFieldCalibration 0s open

Duration by Step

FieldCalibration6m 59sAssignCalFile43s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 43s
FieldCalibration 8 3 2 6m 59s
No action 12 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora142s1_WavelengthCalibration_SlitFuncParam.jpeg - 118 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora142s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 105 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora142s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 91 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora142s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 88 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora142s1_CreateReference_AODEST.jpeg - 47 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora142s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 82 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora142s1_CreateReference_DISPFIT(DISP).jpeg - 83 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora142s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 98 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora142s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 85 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora142s1_FieldCalibration_VC_O3.jpeg - 70 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora142s1_FieldCalibration_SCREFINT_O3.jpeg - 56 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora142s1_FieldCalibration_MLE_NO2.jpeg - 101 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora142s1_FieldCalibration_VCMLE_NO2.jpeg - 81 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora142s1_FieldCalibration_MLE_SO2.jpeg - 63 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora142s1_FieldCalibration_VCMLE_SO2.jpeg - 55 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.737e-04; wrmsmax=1.274e-02; atvarmax(%)=2.000e+01; amflim=(1.300,5.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=12; timeaver(min)=0.00; redu=0; scref=(3.301e-04); uscref=(5.516e-05)

FieldCalibration MLE HCHO

Pandora142s1_FieldCalibration_MLE_HCHO.jpeg - 114 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora142s1_FieldCalibration_VCMLE_HCHO.jpeg - 123 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
looks okay

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
ok

SO2

Decision for SO2
Yes
rather large uncertainty

HCHO

Decision for HCHO
With Care
looks okay, some perioods to low

H2O

Decision for H2O
Yes
okay