Decision Summary

Upload-ready calibration protocol overview
Instrument P141s1
Session 2
Location GreenbeltMD
Calibration person Bryan Place
CF file Pandora141s1_CF_v2d20231011.txt
Software version p1-8
Calibration vector -
Reference day 20231012
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P141s1
Session 2
Events 285
Figures 11
Action time 2h 13m

Action Timeline

86 action intervals
Oct 09, 2023Oct 11, 2023Oct 13, 2023Oct 23, 2023Nov 09, 2023Nov 14, 202308:18 PMCreateReferenceFieldCalibrationAssignCalFileWavelengthCalibrationCreateReference 2m 40s finishedCreateReference 0s abortedCreateReference 34s finishedCreateReference 0s abortedCreateReference 35s finishedCreateReference 0s abortedCreateReference 41s finishedCreateReference 0s abortedCreateReference 31s finishedCreateReference 0s abortedCreateReference 30s finishedCreateReference 0s abortedCreateReference 28s finishedCreateReference 1m 27s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 41s finishedCreateReference 0s abortedCreateReference 1m 21s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 22s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 2s abortedCreateReference 48s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 55s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 20s finishedCreateReference 0s finishedFieldCalibration 30m 32s finishedFieldCalibration 0s abortedFieldCalibration 2s abortedFieldCalibration 5s abortedFieldCalibration 2m 3s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 18s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 5m 50s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 56s abortedFieldCalibration 38m 22s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 14m 19s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 18m 13s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 1m 30s finishedWavelengthCalibration 2m 47s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s staleWavelengthCalibration 2m 12s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s aborted

Duration by Step

FieldCalibration1h 50mCreateReference15m 55sWavelengthCalibration4m 59sAssignCalFile1m 30s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 1m 30s
CreateReference 73 36 13 15m 55s
FieldCalibration 152 41 7 1h 50m
No action 21 0 0 0s
WavelengthCalibration 36 7 1 4m 59s

Session Figures

11 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora141s1_WavelengthCalibration_SlitFuncParam.jpeg - 99 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora141s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 79 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora141s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 60 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora141s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 62 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora141s1_CreateReference_AODEST.jpeg - 26 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora141s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 59 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora141s1_CreateReference_DISPFIT(DISP).jpeg - 60 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora141s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 101 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora141s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 63 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora141s1_FieldCalibration_VC_O3.jpeg - 71 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora141s1_FieldCalibration_SCREFINT_O3.jpeg - 55 KB
FieldCalibration SCREFINT O3

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Calibrated using AXC. All O3 products look good, strong shift in rms observed following change in resolution with temperature

O3Synt

Decision for O3Synt

NO2

Decision for NO2
Calibrated with REF method (P2). Columns look good

SO2

Decision for SO2
Calibrated with REF method (P2). Columns look good and SO2 plume spike capture on 20231105 matches with ref instrument

HCHO

Decision for HCHO
Calibrated with REF method (P2). Total and sky columns look good, but become flagged out due to strong change in resolution and wrms with temperature

H2O

Decision for H2O
H20 columns look good