Decision Summary

Upload-ready calibration protocol overview
Instrument P140s1
Session 10
Location -
Calibration person Manuel Roca
CF file Pandora140s1_CF_v10d20200806.txt
Software version 1.8.49
Calibration vector -
Reference day -
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: YesH2O: Yes
Instrument P140s1
Session 10
Events 19
Figures 17
Action time 29m 0s

Action Timeline

7 action intervals
Mar 13, 2023Mar 14, 202304:06 PM08:44 AMFieldCalibrationAssignCalFileFieldCalibration 24m 51s abortedFieldCalibration 3m 11s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 58s finished

Duration by Step

FieldCalibration28m 2sAssignCalFile58s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 58s
FieldCalibration 10 6 1 28m 2s
No action 6 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora140s1_WavelengthCalibration_SlitFuncParam.jpeg - 241 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora140s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 149 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora140s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 118 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora140s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 181 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora140s1_CreateReference_AODEST.jpeg - 55 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora140s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 155 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora140s1_CreateReference_DISPFIT(DISP).jpeg - 147 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora140s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 283 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora140s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 215 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora140s1_FieldCalibration_VC_O3.jpeg - 160 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora140s1_FieldCalibration_SCREFINT_O3.jpeg - 115 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora140s1_FieldCalibration_MLE_NO2.jpeg - 309 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora140s1_FieldCalibration_VCMLE_NO2.jpeg - 171 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.289e-02; wrmsmax=3.874e-01; atvarmax(%)=2.000e+01; amflim=(1.220,2.000); amflimvar=(0.000,2.000); perc=0.00; ngroup=18; timeaver(min)=0.00; redu=0; scref=(4.280e-05); uscref=(1.683e-05)

FieldCalibration MLE SO2

Pandora140s1_FieldCalibration_MLE_SO2.jpeg - 179 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora140s1_FieldCalibration_VCMLE_SO2.jpeg - 154 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora140s1_FieldCalibration_MLE_HCHO.jpeg - 162 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora140s1_FieldCalibration_VCMLE_HCHO.jpeg - 190 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
looks fine

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
looks good

SO2

Decision for SO2
Yes
looks fine

HCHO

Decision for HCHO
Yes
VC Values too close to climatological in period Mar/Apr-2022

H2O

Decision for H2O
Yes