Decision Summary

Upload-ready calibration protocol overview
Instrument P135s1
Session 15
Location -
Calibration person Stefanie Morhenn
CF file Pandora135s1_CF_v15d20200411.txt
Software version 1.8.52
Calibration vector calV1
Reference day -
Overall status development
O3Ext: With CareO3Synt: YesNO2: YesSO2: With CareHCHO: With CareH2O: Yes
Instrument P135s1
Session 15
Events 84
Figures 17
Action time 4h 40m

Action Timeline

36 action intervals
Mar 12, 2023Mar 28, 202311:18 AM06:06 PMFieldCalibrationAssignCalFileFieldCalibration 12s abortedFieldCalibration 13s abortedFieldCalibration 25s abortedFieldCalibration 1m 42s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 23s abortedFieldCalibration 3h 34m abortedFieldCalibration 1m 8s abortedFieldCalibration 1m 19s abortedFieldCalibration 1m 33s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 4m 2s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 4s abortedFieldCalibration 12s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedAssignCalFile 25s finishedFieldCalibration 46m 58s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 6m 29s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 45s finished

Duration by Step

FieldCalibration4h 39mAssignCalFile1m 10s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 6 2 0 1m 10s
FieldCalibration 57 34 11 4h 39m
No action 21 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora135s1_WavelengthCalibration_SlitFuncParam.jpeg - 239 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora135s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 231 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora135s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 199 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora135s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 193 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora135s1_CreateReference_AODEST.jpeg - 84 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora135s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 162 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora135s1_CreateReference_DISPFIT(DISP).jpeg - 154 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora135s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 172 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora135s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 168 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora135s1_FieldCalibration_VC_O3.jpeg - 114 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora135s1_FieldCalibration_SCREFINT_O3.jpeg - 73 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 redu=1 settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.137e-04; wrmsmax=8.000e-04; atvarmax(%)=3.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=24; timeaver(min)=0.00; redu=1; scref=(1.690e-04,1.395e-04,1.465e-04); uscref=(1.325e-05,8.143e-06,1.285e-05)

FieldCalibration MLE NO2

Pandora135s1_FieldCalibration_MLE_NO2.jpeg - 234 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora135s1_FieldCalibration_VCMLE_NO2.jpeg - 108 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora135s1_FieldCalibration_MLE_SO2.jpeg - 184 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora135s1_FieldCalibration_VCMLE_SO2.jpeg - 190 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO redu=1 settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.176e-04; wrmsmax=8.000e-04; atvarmax(%)=3.000e+01; amflim=(1.400,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=1; scref=(2.456e-04); uscref=(3.107e-05)

FieldCalibration MLE HCHO

Pandora135s1_FieldCalibration_MLE_HCHO.jpeg - 120 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora135s1_FieldCalibration_VCMLE_HCHO.jpeg - 108 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
AXC looked good, unc=7.62e-04

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
okay, E-MLE used

SO2

Decision for SO2
With Care
sometimes too high (inverse-u), sometimes too low (below zero)

HCHO

Decision for HCHO
With Care
sometimes too high (inverse-u shaped), sometimes too low (below sky data)

H2O

Decision for H2O
Yes
okay