Decision Summary
Upload-ready calibration protocol overview
Instrument
P135s1
Session
15
Location
-
Calibration person
Stefanie Morhenn
CF file
Pandora135s1_CF_v15d20200411.txt
Software version
1.8.52
Calibration vector
calV1
Reference day
-
Overall status
development
O3Ext: With CareO3Synt: YesNO2: YesSO2: With CareHCHO: With CareH2O: Yes
Action Timeline
36 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 6 | 2 | 0 | 1m 10s |
| FieldCalibration | 57 | 34 | 11 | 4h 39m |
| No action | 21 | 0 | 0 | 0s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora135s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 231 KBCreate reference SELSPECS
Pandora135s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 199 KBCreate reference SLERR (STRAY)
Pandora135s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 193 KBCreate reference AODEST
Pandora135s1_CreateReference_AODEST.jpeg - 84 KBCreate reference SENSCORRFIN (SENS)
Pandora135s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 162 KBCreate reference DISPFIT (DISP)
Pandora135s1_CreateReference_DISPFIT(DISP).jpeg - 154 KBCreate reference WLCHANGE (DISP)
Pandora135s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 172 KBCreate reference SHIFT (CLOS)
Pandora135s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 168 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
AXC looked good, unc=7.62e-04
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
Yes
okay, E-MLE used
SO2
Decision for SO2
With Care
sometimes too high (inverse-u), sometimes too low (below zero)
HCHO
Decision for HCHO
With Care
sometimes too high (inverse-u shaped), sometimes too low (below sky data)
H2O
Decision for H2O
Yes
okay