Decision Summary

Upload-ready calibration protocol overview
Instrument P133s1
Session 4
Location Heidelberg
Calibration person Markus Kilian
CF file Pandora133s1_CF_v4d20240611.txt
Software version 1.8.75
Calibration vector calV1
Reference day 20240826
Overall status development
O3Ext: YesO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Yes
Instrument P133s1
Session 4
Events 418
Figures 17
Action time 1h 28m

Action Timeline

97 action intervals
Jul 04, 202510:19 AM01:06 PMCreateReferenceFieldCalibrationAssignCalFileCreateReference 18s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 15s finishedCreateReference 0s abortedCreateReference 1m 1s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 17s abortedCreateReference 1m 11s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 14s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 58s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 16s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 13s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 12s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 19s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 20s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 16s finishedCreateReference 0s finishedFieldCalibration 6m 42s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 29s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 40m 16s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 1m 1s abortedFieldCalibration 2m 8s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 2m 14s abortedFieldCalibration 20s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 1m 3s abortedFieldCalibration 13s abortedFieldCalibration 5m 6s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 1m 8s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 4m 54s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 11s abortedFieldCalibration 7s abortedFieldCalibration 8s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 16m 29s finished

Duration by Step

FieldCalibration1h 6mAssignCalFile16m 29sCreateReference5m 50s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 16m 29s
CreateReference 87 41 12 5m 50s
FieldCalibration 313 55 12 1h 6m
No action 15 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora133s1_WavelengthCalibration_SlitFuncParam.jpeg - 200 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora133s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 105 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora133s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 85 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora133s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 84 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora133s1_CreateReference_AODEST.jpeg - 48 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora133s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 90 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora133s1_CreateReference_DISPFIT(DISP).jpeg - 83 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora133s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 80 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora133s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 104 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora133s1_FieldCalibration_VC_O3.jpeg - 69 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora133s1_FieldCalibration_SCREFINT_O3.jpeg - 54 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.272e-03; wrmsmax=9.000e-04; atvarmax(%)=6.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=26; timeaver(min)=0.00; redu=2; scref=(1.339e-04,1.171e-04,1.220e-04); uscref=(1.274e-05,1.074e-05,1.455e-05)

FieldCalibration MLE NO2

Pandora133s1_FieldCalibration_MLE_NO2.jpeg - 96 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora133s1_FieldCalibration_VCMLE_NO2.jpeg - 73 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.577e-02; wrmsmax=9.000e-04; atvarmax(%)=6.000e+01; amflim=(1.000,2.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(1.072e-05); uscref=(2.611e-05)

FieldCalibration MLE SO2

Pandora133s1_FieldCalibration_MLE_SO2.jpeg - 68 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora133s1_FieldCalibration_VCMLE_SO2.jpeg - 62 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=9.076e-03; wrmsmax=1.969e-01; atvarmax(%)=6.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(9.854e-05); uscref=(1.154e-05)

FieldCalibration MLE HCHO

Pandora133s1_FieldCalibration_MLE_HCHO.jpeg - 60 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora133s1_FieldCalibration_VCMLE_HCHO.jpeg - 59 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
O3Synt bad in winter and flagged out.

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
E-MLE

SO2

Decision for SO2
Yes
ok; but flagged out in winter months cause of FOV problem of U340 Products.

HCHO

Decision for HCHO
Yes
ok, E-MLE used; flagged out in winter months cause of FOV problem of U340 Products.

H2O

Decision for H2O
Yes