Decision Summary
Upload-ready calibration protocol overview
Instrument
P132s1
Session
15
Location
Berlin
Calibration person
Stefanie Morhenn
CF file
Pandora132s1_CF_v15d20171113.txt
Software version
1.8.88
Calibration vector
calV3
Reference day
20180429
Overall status
development
O3Ext: UnusableO3Synt: UnusableNO2: YesSO2: UnusableHCHO: UnusableH2O: Yes
Action Timeline
150 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 1m 16s |
| Convolution | 36 | 12 | 2 | 2m 37s |
| CreateReference | 161 | 73 | 23 | 26m 20s |
| FieldCalibration | 273 | 63 | 13 | 9h 50m |
| No action | 16 | 0 | 0 | 0s |
| SensitivityCalibration | 2 | 1 | 1 | 2s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora132s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 149 KBCreate reference SELSPECS
Pandora132s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 129 KBCreate reference SLERR (STRAY)
Pandora132s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 76 KBCreate reference AODEST
Pandora132s1_CreateReference_AODEST.jpeg - 64 KBCreate reference SENSCORRFIN (SENS)
Pandora132s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 101 KBCreate reference DISPFIT (DISP)
Pandora132s1_CreateReference_DISPFIT(DISP).jpeg - 103 KBCreate reference WLCHANGE (DISP)
Pandora132s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 105 KBCreate reference SHIFT (CLOS)
Pandora132s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 178 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Unusable
wrong fiber connected in the field - U340 products unusable
O3Synt
Decision for O3Synt
Unusable
NO2
Decision for NO2
Yes
looks good
SO2
Decision for SO2
Unusable
wrong fiber connected in the field - U340 products unusable
HCHO
Decision for HCHO
Unusable
wrong fiber connected in the field - U340 products unusable
H2O
Decision for H2O
Yes