Decision Summary
Upload-ready calibration protocol overview
Instrument
P132s1
Session
13
Location
Berlin
Calibration person
Markus Kilian
CF file
Pandora132s1_CF_v3d20210422.txt
Software version
p1-8
Calibration vector
-
Reference day
20210427
Overall status
development
O3Ext: YesO3Synt: YesNO2: YesSO2: With CareHCHO: YesH2O: Yes
Action Timeline
71 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 17s |
| CreateReference | 75 | 38 | 13 | 33m 50s |
| FieldCalibration | 211 | 32 | 0 | 51m 47s |
| No action | 15 | 0 | 0 | 0s |
Session Figures
No session figures found for the loaded instrument/session.No figures loaded
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
ous1 and O3T have wiggly wrms
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
Yes
EMLE; wiggly wrms
SO2
Decision for SO2
With Care
MLE
HCHO
Decision for HCHO
Yes
EMLE; wiggly wrms
H2O
Decision for H2O
Yes