Decision Summary

Upload-ready calibration protocol overview
Instrument P132s1
Session 13
Location Berlin
Calibration person Markus Kilian
CF file Pandora132s1_CF_v3d20210422.txt
Software version p1-8
Calibration vector -
Reference day 20210427
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: With CareHCHO: YesH2O: Yes
Instrument P132s1
Session 13
Events 304
Figures 0
Action time 1h 25m

Action Timeline

71 action intervals
Sep 19, 2024Sep 23, 2024Nov 15, 202406:53 AM02:12 PMCreateReferenceFieldCalibrationAssignCalFileCreateReference 24s finishedCreateReference 0s abortedCreateReference 18s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 23s finishedCreateReference 0s abortedCreateReference 1m 43s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 14s finishedCreateReference 0s abortedCreateReference 15m 18s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 20s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 8s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 7m 20s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 28s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 39s finishedCreateReference 0s finishedFieldCalibration 1m 53s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 33m 36s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 5m 1s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 11m 17s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 17s finishedCreateReference 21s finishedCreateReference 0s abortedCreateReference 5m 5s abortedCreateReference 9s aborted

Duration by Step

FieldCalibration51m 47sCreateReference33m 50sAssignCalFile17s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 17s
CreateReference 75 38 13 33m 50s
FieldCalibration 211 32 0 51m 47s
No action 15 0 0 0s

Session Figures

No session figures found for the loaded instrument/session.
No figures loaded

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
ous1 and O3T have wiggly wrms

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
EMLE; wiggly wrms

SO2

Decision for SO2
With Care
MLE

HCHO

Decision for HCHO
Yes
EMLE; wiggly wrms

H2O

Decision for H2O
Yes