Decision Summary
Upload-ready calibration protocol overview
Instrument
P130s1
Session
4
Location
LabSciGlob
Calibration person
Ragi Rajagopalan
CF file
Pandora130s1_CF_v4d20190527.txt
Software version
1.8.49
Calibration vector
-
Reference day
20190624
Overall status
development
O3Ext: UnusableO3Synt: With CareNO2: With CareSO2: YesHCHO: YesH2O: Not answered
Action Timeline
311 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 21 | 3 | 2 | 13m 17s |
| AnalyzeDark | 25 | 5 | 4 | 16m 12s |
| AssignCalFile | 3 | 1 | 0 | 1m 17s |
| CheckFilters | 6 | 2 | 1 | 57s |
| CheckNoiseWave | 8 | 2 | 1 | 1m 9s |
| Convolution | 68 | 22 | 2 | 5m 27s |
| CreateReference | 203 | 101 | 38 | 24m 9s |
| FieldCalibration | 363 | 93 | 28 | 2h 32m |
| NDFilters | 741 | 27 | 18 | 48m 21s |
| No action | 31 | 0 | 0 | 0s |
| SensitivityCalibration | 802 | 39 | 19 | 57m 5s |
| StrayLight | 6 | 2 | 2 | 19s |
| TempDependence | 11 | 3 | 2 | 1m 32s |
| WavelengthCalibration | 32 | 10 | 2 | 7m 42s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora130s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 159 KBCreate reference SELSPECS
Pandora130s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 133 KBCreate reference SLERR (STRAY)
Pandora130s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 117 KBCreate reference AODEST
Pandora130s1_CreateReference_AODEST.jpeg - 56 KBCreate reference SENSCORRFIN (SENS)
Pandora130s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 127 KBCreate reference DISPFIT (DISP)
Pandora130s1_CreateReference_DISPFIT(DISP).jpeg - 98 KBCreate reference WLCHANGE (DISP)
Pandora130s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 205 KBCreate reference SHIFT (CLOS)
Pandora130s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 133 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Unusable
was working fine until 2019-10-10 and after that seems to have spectrometer connectivity issues,spectrometer started working again by itself from 2020-03-13 but data was flagged out due to high wrms.
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
With Care
Looks fine until 2019-10-10.
SO2
Decision for SO2
Yes
Looks fine until 2019-10-10
HCHO
Decision for HCHO
Yes
Looks fine until 2019-10-10
H2O
Decision for H2O