Decision Summary

Upload-ready calibration protocol overview
Instrument P128s1
Session 7
Location -
Calibration person BPlace
CF file -
Software version 1.8.60
Calibration vector -
Reference day -
Overall status development
O3Ext: UnusableO3Synt: With CareNO2: YesSO2: YesHCHO: With CareH2O: Not answered
Instrument P128s1
Session 7
Events 0
Figures 0
Action time 0s

Action Timeline

No start/finish intervals found.
No timeline intervals found

Duration by Step

No completed durations found

Action Summary

Action Events Runs Aborted Time
No actions loaded

Session Figures

No session figures found for the loaded instrument/session.
No figures loaded

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable
Calibrated with REF method against P2. O3T estimated as >10K from climatology - high stray light unit. O3Synt product looks good during clearance period but may drift with resolution change

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
Calibrated with REF method against P2, NO2 products look good

SO2

Decision for SO2
Yes
Calibrated with REF method against P2, SO2 looks good

HCHO

Decision for HCHO
With Care
Calibrated with REF method against P2, columns and rms look good but unit has resolution change drift that may flag HCHO out

H2O

Decision for H2O