Decision Summary

Upload-ready calibration protocol overview
Instrument P127s1
Session 7
Location -
Calibration person Markus Kilian
CF file Pandora127s1_CF_v7d20250721.txt
Software version 1.8.75
Calibration vector calV2
Reference day -
Overall status development
O3Ext: YesO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Yes
Instrument P127s1
Session 7
Events 20
Figures 17
Action time 15m 12s

Action Timeline

6 action intervals
Oct 06, 202501:38 PM02:14 PMFieldCalibrationAssignCalFileFieldCalibration 13m 30s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 1m 42s finished

Duration by Step

FieldCalibration13m 30sAssignCalFile1m 42s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 1m 42s
FieldCalibration 8 5 0 13m 30s
No action 9 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora127s1_WavelengthCalibration_SlitFuncParam.jpeg - 141 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora127s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 117 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora127s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 104 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora127s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 114 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora127s1_CreateReference_AODEST.jpeg - 46 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora127s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 90 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora127s1_CreateReference_DISPFIT(DISP).jpeg - 77 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora127s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 86 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora127s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 109 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora127s1_FieldCalibration_VC_O3.jpeg - 70 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora127s1_FieldCalibration_SCREFINT_O3.jpeg - 54 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora127s1_FieldCalibration_MLE_NO2.jpeg - 80 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora127s1_FieldCalibration_VCMLE_NO2.jpeg - 80 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.503e-02; wrmsmax=9.000e-04; atvarmax(%)=2.000e+01; amflim=(1.284,2.400); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(4.905e-05); uscref=(5.500e-05)

FieldCalibration MLE SO2

Pandora127s1_FieldCalibration_MLE_SO2.jpeg - 64 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora127s1_FieldCalibration_VCMLE_SO2.jpeg - 62 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora127s1_FieldCalibration_MLE_HCHO.jpeg - 71 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora127s1_FieldCalibration_VCMLE_HCHO.jpeg - 81 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
O3Ext mostly flagged out, O3Synt looks good.

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
E-MLE used, data looks good.

SO2

Decision for SO2
Yes
MLE correction by lowering SC in comparison to session 5 and CF v6.

HCHO

Decision for HCHO
Yes
E-MLE used, data looks good.

H2O

Decision for H2O
Yes