Decision Summary
Upload-ready calibration protocol overview
Instrument
P127s1
Session
5
Location
Mainz-MPIC
Calibration person
Markus Kilian
CF file
Pandora127s1_CF_v5d20250721.txt
Software version
1.8.75
Calibration vector
-
Reference day
20250814
Overall status
development
O3Ext: YesO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Yes
Action Timeline
165 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 3m 51s |
| CheckFilters | 3 | 1 | 0 | 15s |
| CreateReference | 223 | 106 | 38 | 40m 38s |
| FieldCalibration | 195 | 54 | 20 | 3h 3m |
| NDFilters | 2 | 1 | 1 | 8s |
| No action | 23 | 0 | 0 | 0s |
| WavelengthCalibration | 5 | 2 | 1 | 3m 4s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora127s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 117 KBCreate reference SELSPECS
Pandora127s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 104 KBCreate reference SLERR (STRAY)
Pandora127s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 114 KBCreate reference AODEST
Pandora127s1_CreateReference_AODEST.jpeg - 46 KBCreate reference SENSCORRFIN (SENS)
Pandora127s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 90 KBCreate reference DISPFIT (DISP)
Pandora127s1_CreateReference_DISPFIT(DISP).jpeg - 77 KBCreate reference WLCHANGE (DISP)
Pandora127s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 86 KBCreate reference SHIFT (CLOS)
Pandora127s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 109 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
O3Ext mostly flagged out, O3Synt looks good.
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
E-MLE used, data looks good.
SO2
Decision for SO2
Yes
okay, but rather high uncertainty
HCHO
Decision for HCHO
Yes
E-MLE used, data looks good.
H2O
Decision for H2O
Yes