Decision Summary

Upload-ready calibration protocol overview
Instrument P124s1
Session 3
Location ComodoroRivadavia
Calibration person Manuel Gebetsberger
CF file Pandora124s1_CF_v1d20180726.txt
Software version 1.8.22
Calibration vector -
Reference day -
Overall status development
O3Ext: YesO3Synt: UnusableNO2: YesSO2: With CareHCHO: UnusableH2O: Not answered
Instrument P124s1
Session 3
Events 103
Figures 15
Action time 1h 2m

Action Timeline

22 action intervals
Nov 26, 2019Jan 29, 2020Jan 30, 2020Feb 11, 202011:19 AM12:30 PMCreateReferenceCheckDispersionAssignCalFileCreateReference 0s staleCheckDispersion 8m 0s abortedCheckDispersion 3m 13s finishedCheckDispersion 1m 55s finishedCreateReference 9m 6s abortedCheckDispersion 6m 17s finishedCheckDispersion 50s finishedCheckDispersion 0s finishedCheckDispersion 2m 23s finishedCreateReference 5m 9s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 8m 35s abortedCreateReference 3m 4s abortedCreateReference 2m 49s abortedCreateReference 9m 39s finishedCreateReference 0s finishedCreateReference 0s abortedAssignCalFile 51s finishedCreateReference 47s aborted

Duration by Step

CreateReference39m 8sCheckDispersion22m 38sAssignCalFile51s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 51s
CheckDispersion 24 7 1 22m 38s
CreateReference 66 13 7 39m 8s
No action 10 0 0 0s

Session Figures

15 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora124s1_WavelengthCalibration_SlitFuncParam.jpeg - 112 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora124s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 112 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora124s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 96 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora124s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 84 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora124s1_CreateReference_AODEST.jpeg - 38 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora124s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 93 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora124s1_CreateReference_DISPFIT(DISP).jpeg - 68 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora124s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 94 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora124s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 112 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora124s1_FieldCalibration_VC_O3.jpeg - 92 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora124s1_FieldCalibration_SCREFINT_O3.jpeg - 53 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora124s1_FieldCalibration_MLE_NO2.jpeg - 119 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora124s1_FieldCalibration_VCMLE_NO2.jpeg - 118 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora124s1_FieldCalibration_MLE_SO2.jpeg - 83 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora124s1_FieldCalibration_VCMLE_SO2.jpeg - 74 KB
FieldCalibration VCMLE SO2

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
probably bias due into missing SLC in reference, affecting daily shape which is slightly U

O3Synt

Decision for O3Synt
Unusable

NO2

Decision for NO2
Yes
ok

SO2

Decision for SO2
With Care
AMF limitation by 1.5, high uncertainties in the measurements

HCHO

Decision for HCHO
Unusable
not calibrated because of Delrin

H2O

Decision for H2O