Decision Summary

Upload-ready calibration protocol overview
Instrument P124s1
Session 2
Location ComodoroRivadavia
Calibration person Manuel Gebetsberger
CF file Pandora124s1_CF_v1d20170602.txt
Software version 1.8.22
Calibration vector -
Reference day -
Overall status development
O3Ext: YesO3Synt: UnusableNO2: YesSO2: With CareHCHO: UnusableH2O: Not answered
Instrument P124s1
Session 2
Events 101
Figures 15
Action time 7h 14m

Action Timeline

17 action intervals
Jan 28, 2020Jan 29, 202001:17 PM08:55 AMCreateReferenceCheckDispersionAssignCalFileCreateReference 16s abortedCreateReference 6h 44m abortedCheckDispersion 3m 42s finishedCheckDispersion 10s finishedCheckDispersion 0s finishedCreateReference 4m 37s abortedCreateReference 11s abortedCreateReference 40s abortedCreateReference 4m 36s abortedCreateReference 15m 41s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedAssignCalFile 43s finished

Duration by Step

CreateReference7h 10mCheckDispersion3m 52sAssignCalFile43s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 4 1 0 43s
CheckDispersion 10 3 0 3m 52s
CreateReference 83 13 7 7h 10m
No action 4 0 0 0s

Session Figures

15 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora124s1_WavelengthCalibration_SlitFuncParam.jpeg - 112 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora124s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 108 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora124s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 89 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora124s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 84 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora124s1_CreateReference_AODEST.jpeg - 38 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora124s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 88 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora124s1_CreateReference_DISPFIT(DISP).jpeg - 72 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora124s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 135 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora124s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 125 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora124s1_FieldCalibration_VC_O3.jpeg - 85 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora124s1_FieldCalibration_SCREFINT_O3.jpeg - 53 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora124s1_FieldCalibration_MLE_NO2.jpeg - 119 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora124s1_FieldCalibration_VCMLE_NO2.jpeg - 76 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora124s1_FieldCalibration_MLE_SO2.jpeg - 93 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora124s1_FieldCalibration_VCMLE_SO2.jpeg - 97 KB
FieldCalibration VCMLE SO2

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
extrapolation, but uncertainty still okay at 1.579155e-03

O3Synt

Decision for O3Synt
Unusable

NO2

Decision for NO2
Yes
okay

SO2

Decision for SO2
With Care
AMF limitation by 1.5, high uncertainties in the measurements

HCHO

Decision for HCHO
Unusable
not calibrated because of Delrin

H2O

Decision for H2O