Decision Summary

Upload-ready calibration protocol overview
Instrument P122s1
Session 10
Location Downsview
Calibration person Christoph Waldauf
CF file Pandora122s1_CF_v1d20220801.txt
Software version 1.8.58
Calibration vector calV2
Reference day -
Overall status development
O3Ext: UnusableO3Synt: UnusableNO2: YesSO2: UnusableHCHO: UnusableH2O: Yes
Instrument P122s1
Session 10
Events 263
Figures 11
Action time 1h 57m

Action Timeline

32 action intervals
Mar 16, 2026Mar 17, 2026Mar 31, 202602:39 PMFieldCalibrationCreateReferenceAssignCalFileFieldCalibration 1m 41s abortedCreateReference 23s abortedFieldCalibration 15s abortedFieldCalibration 21s abortedFieldCalibration 0s staleFieldCalibration 0s staleFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 14m 32s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 1h 25m finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 10m 47s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedAssignCalFile 1m 6s finishedCreateReference 47s abortedCreateReference 11s abortedCreateReference 24s abortedCreateReference 1m 28s aborted

Duration by Step

FieldCalibration1h 53mCreateReference3m 12sAssignCalFile1m 6s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 1m 6s
CreateReference 10 5 5 3m 12s
FieldCalibration 241 24 7 1h 53m
No action 9 0 0 0s

Session Figures

11 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora122s1_WavelengthCalibration_SlitFuncParam.jpeg - 244 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora122s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 241 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora122s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 168 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora122s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 180 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora122s1_CreateReference_AODEST.jpeg - 91 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora122s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 158 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora122s1_CreateReference_DISPFIT(DISP).jpeg - 166 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora122s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 176 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora122s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 178 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora122s1_FieldCalibration_VC_O3.jpeg - 143 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora122s1_FieldCalibration_SCREFINT_O3.jpeg - 105 KB
FieldCalibration SCREFINT O3

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable
switched fibers

O3Synt

Decision for O3Synt
Unusable

NO2

Decision for NO2
Yes
looks ok

SO2

Decision for SO2
Unusable
switched fibers

HCHO

Decision for HCHO
Unusable
swithed fibers/Delrin

H2O

Decision for H2O
Yes
looks ok