Decision Summary

Upload-ready calibration protocol overview
Instrument P121s1
Session 59
Location Izana
Calibration person Ragi Rajagopalan
CF file Pandora121s1_CF_v59d20181203.txt
Software version 1.8.73
Calibration vector calV2
Reference day 20190508
Overall status development
O3Ext: UnusableO3Synt: YesNO2: YesSO2: With CareHCHO: UnusableH2O: Yes
Instrument P121s1
Session 59
Events 102
Figures 17
Action time 1h 12m

Action Timeline

43 action intervals
Feb 05, 202610:14 AM01:46 PMCreateReferenceFieldCalibrationAssignCalFileCreateReference 18m 56s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 23s finishedCreateReference 0s abortedCreateReference 19s abortedCreateReference 14s abortedCreateReference 32s abortedCreateReference 40s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 3m 25s finishedCreateReference 0s abortedCreateReference 1m 11s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 40s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 27s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 23s abortedCreateReference 41m 22s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 17s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 15s finishedCreateReference 0s finishedFieldCalibration 14s finishedFieldCalibration 0s abortedCreateReference 2s abortedFieldCalibration 11s abortedFieldCalibration 2m 27s finishedFieldCalibration 0s abortedAssignCalFile 29s finished

Duration by Step

CreateReference1h 9mFieldCalibration2m 53sAssignCalFile29s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 29s
CreateReference 75 37 14 1h 9m
FieldCalibration 13 5 3 2m 53s
No action 11 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora121s1_WavelengthCalibration_SlitFuncParam.jpeg - 165 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora121s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 143 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora121s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 121 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora121s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 137 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora121s1_CreateReference_AODEST.jpeg - 64 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora121s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 105 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora121s1_CreateReference_DISPFIT(DISP).jpeg - 121 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora121s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 133 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora121s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 125 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora121s1_FieldCalibration_VC_O3.jpeg - 101 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora121s1_FieldCalibration_SCREFINT_O3.jpeg - 75 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora121s1_FieldCalibration_MLE_NO2.jpeg - 149 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora121s1_FieldCalibration_VCMLE_NO2.jpeg - 99 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.818e-02; wrmsmax=1.058e+00; atvarmax(%)=1.000e+02; amflim=(1.220,3.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=15; timeaver(min)=0.00; redu=0; scref=(5.849e-05); uscref=(1.395e-05)

FieldCalibration MLE SO2

Pandora121s1_FieldCalibration_MLE_SO2.jpeg - 113 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora121s1_FieldCalibration_VCMLE_SO2.jpeg - 113 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora121s1_FieldCalibration_MLE_HCHO.jpeg - 68 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora121s1_FieldCalibration_VCMLE_HCHO.jpeg - 81 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes

SO2

Decision for SO2
With Care
according to new norms range within 0 to 1e-4 but has the werid AMF/O3T dependent shapes

HCHO

Decision for HCHO
Unusable

H2O

Decision for H2O
Yes
A bit higher than P209 but looks good