Decision Summary

Upload-ready calibration protocol overview
Instrument P121s1
Session 51
Location -
Calibration person Ragi Rajagopalan
CF file C:/Blick_git/data/calresults/Pandora121s1/session51/Pandora121s1_CF_session51.txt
Software version 1.8.57
Calibration vector -
Reference day -
Overall status development
O3Ext: UnusableO3Synt: YesNO2: YesSO2: With CareHCHO: UnusableH2O: Yes
Instrument P121s1
Session 51
Events 9
Figures 17
Action time 0s

Action Timeline

No start/finish intervals found.
No timeline intervals found

Duration by Step

No completed durations found

Action Summary

Action Events Runs Aborted Time
No action 9 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora121s1_WavelengthCalibration_SlitFuncParam.jpeg - 244 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora121s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 232 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora121s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 187 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora121s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 177 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora121s1_CreateReference_AODEST.jpeg - 84 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora121s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 174 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora121s1_CreateReference_DISPFIT(DISP).jpeg - 153 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora121s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 289 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora121s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 184 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora121s1_FieldCalibration_VC_O3.jpeg - 110 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora121s1_FieldCalibration_SCREFINT_O3.jpeg - 80 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora121s1_FieldCalibration_MLE_NO2.jpeg - 195 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora121s1_FieldCalibration_VCMLE_NO2.jpeg - 137 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora121s1_FieldCalibration_MLE_SO2.jpeg - 202 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora121s1_FieldCalibration_VCMLE_SO2.jpeg - 220 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora121s1_FieldCalibration_MLE_HCHO.jpeg - 184 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora121s1_FieldCalibration_VCMLE_HCHO.jpeg - 133 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable
O3T is high most times and out2 is flagged out most of the time

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
MLE used, EMLE values were less than startospheric

SO2

Decision for SO2
With Care
ok

HCHO

Decision for HCHO
Unusable
daily sudden jumps recommended not to use

H2O

Decision for H2O
Yes