Decision Summary

Upload-ready calibration protocol overview
Instrument P120s1
Session 7
Location -
Calibration person Ragi Rajagopalan
CF file Pandora120s1_CF_v7d20240814.txt
Software version 1.8.92
Calibration vector calV2
Reference day -
Overall status development
O3Ext: YesO3Synt: With CareNO2: YesSO2: YesHCHO: UnusableH2O: Yes
Instrument P120s1
Session 7
Events 22
Figures 17
Action time 3m 40s

Action Timeline

5 action intervals
Mar 23, 202609:47 AM10:25 AMFieldCalibrationAssignCalFileFieldCalibration 12s finishedFieldCalibration 0s abortedFieldCalibration 11s abortedFieldCalibration 2m 52s abortedAssignCalFile 26s finished

Duration by Step

FieldCalibration3m 15sAssignCalFile26s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 26s
FieldCalibration 8 4 3 3m 15s
No action 11 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora120s1_WavelengthCalibration_SlitFuncParam.jpeg - 124 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora120s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 190 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora120s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 139 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora120s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 131 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora120s1_CreateReference_AODEST.jpeg - 76 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora120s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 133 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora120s1_CreateReference_DISPFIT(DISP).jpeg - 140 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora120s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 129 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora120s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 166 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora120s1_FieldCalibration_VC_O3.jpeg - 159 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora120s1_FieldCalibration_SCREFINT_O3.jpeg - 89 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora120s1_FieldCalibration_MLE_NO2.jpeg - 187 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora120s1_FieldCalibration_VCMLE_NO2.jpeg - 113 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=8.871e-04; wrmsmax=5.446e-02; atvarmax(%)=5.000e+01; amflim=(1.000,3.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=7; timeaver(min)=0.00; redu=0; scref=(4.491e-05); uscref=(4.924e-05)

FieldCalibration MLE SO2

Pandora120s1_FieldCalibration_MLE_SO2.jpeg - 55 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora120s1_FieldCalibration_VCMLE_SO2.jpeg - 49 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora120s1_FieldCalibration_MLE_HCHO.jpeg - 120 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora120s1_FieldCalibration_VCMLE_HCHO.jpeg - 123 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
O3 Temp has a spread

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes

SO2

Decision for SO2
Yes

HCHO

Decision for HCHO
Unusable
specres.hcho;broad band structure leading to negative VC plus fine scale structure showing jumpging vertical columns

H2O

Decision for H2O