Decision Summary
Upload-ready calibration protocol overview
Instrument
P120s1
Session
3
Location
IBK
Calibration person
Manuel Gebetsberger
CF file
Pandora120s1_CF_v1d20160404.txt
Software version
1.8.33
Calibration vector
-
Reference day
20161206
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: With CareHCHO: UnusableH2O: Not answered
Action Timeline
112 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 16 | 2 | 1 | 3m 42s |
| AnalyzeDark | 4 | 2 | 2 | 16s |
| AssignCalFile | 3 | 1 | 0 | 1m 37s |
| CheckFilters | 4 | 1 | 0 | 1m 16s |
| CheckNoiseWave | 6 | 1 | 0 | 24s |
| Convolution | 18 | 6 | 1 | 50s |
| CreateReference | 102 | 47 | 17 | 12m 13s |
| FieldCalibration | 155 | 41 | 11 | 1h 18m |
| NDFilters | 136 | 3 | 1 | 7m 58s |
| No action | 18 | 0 | 0 | 0s |
| SensitivityCalibration | 220 | 8 | 2 | 5m 1s |
Session Figures
15 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora120s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 115 KBCreate reference SELSPECS
Pandora120s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 88 KBCreate reference SLERR (STRAY)
Pandora120s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 80 KBCreate reference AODEST
Pandora120s1_CreateReference_AODEST.jpeg - 38 KBCreate reference SENSCORRFIN (SENS)
Pandora120s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 93 KBCreate reference DISPFIT (DISP)
Pandora120s1_CreateReference_DISPFIT(DISP).jpeg - 76 KBCreate reference WLCHANGE (DISP)
Pandora120s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 128 KBCreate reference SHIFT (CLOS)
Pandora120s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 122 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
MLE with redu=1, not sky data available
SO2
Decision for SO2
With Care
no events
HCHO
Decision for HCHO
Unusable
delrin
H2O
Decision for H2O