Decision Summary
Upload-ready calibration protocol overview
Instrument
P120s1
Session
2
Location
Davos
Calibration person
Manuel Gebetsberger
CF file
Pandora120s1_CF_v1d20200806.txt
Software version
p1-8
Calibration vector
calV1
Reference day
20200908
Overall status
development
O3Ext: YesO3Synt: With CareNO2: YesSO2: YesHCHO: With CareH2O: Yes
Action Timeline
89 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 6 | 2 | 0 | 3m 34s |
| CheckNoiseWave | 3 | 1 | 1 | 3s |
| CreateReference | 67 | 33 | 10 | 12m 41s |
| FieldCalibration | 255 | 51 | 7 | 8h 31m |
| No action | 18 | 0 | 0 | 0s |
| SensitivityCalibration | 2 | 1 | 1 | 25s |
| WavelengthCalibration | 2 | 1 | 1 | 39s |
Session Figures
No session figures found for the loaded instrument/session.No figures loaded
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
okay
SO2
Decision for SO2
Yes
no events visible but a nice flat background
HCHO
Decision for HCHO
With Care
noisy pattern around noon
H2O
Decision for H2O
Yes
very good aggremant with Aeronet