Decision Summary

Upload-ready calibration protocol overview
Instrument P120s1
Session 2
Location Davos
Calibration person Manuel Gebetsberger
CF file Pandora120s1_CF_v1d20200806.txt
Software version p1-8
Calibration vector calV1
Reference day 20200908
Overall status development
O3Ext: YesO3Synt: With CareNO2: YesSO2: YesHCHO: With CareH2O: Yes
Instrument P120s1
Session 2
Events 353
Figures 0
Action time 8h 49m

Action Timeline

89 action intervals
Jun 10, 2022 - Jun 11, 2022Jun 12, 2022Jun 13, 2022Jun 14, 2022Jun 23, 2022Jun 24, 2022Jul 07, 202209:20 PMCreateReferenceFieldCalibrationAssignCalFileCheckNoiseWaveWavelengthCalibrationSensitivityCalibrationCreateReference 52s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 53s finishedCreateReference 0s abortedCreateReference 1m 27s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 28s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 25s finishedCreateReference 0s abortedCreateReference 39s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 7s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 4s abortedCreateReference 1m 40s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 11s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 55s finishedCreateReference 0s finishedFieldCalibration 2m 43s abortedFieldCalibration 1m 24s finishedFieldCalibration 0s abortedFieldCalibration 19s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 7h 30m finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 5s abortedFieldCalibration 1m 8s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 9m 42s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 14m 19s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 2m 56s finishedFieldCalibration 6s abortedFieldCalibration 5s abortedFieldCalibration 19m 1s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 4m 54s finishedFieldCalibration 0s finishedAssignCalFile 38s finishedFieldCalibration 7m 20s abortedCheckNoiseWave 3s abortedWavelengthCalibration 39s abortedSensitivityCalibration 25s aborted

Duration by Step

FieldCalibration8h 31mCreateReference12m 41sAssignCalFile3m 34sWavelengthCalibration39sSensitivityCalibration25sCheckNoiseWave3s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 6 2 0 3m 34s
CheckNoiseWave 3 1 1 3s
CreateReference 67 33 10 12m 41s
FieldCalibration 255 51 7 8h 31m
No action 18 0 0 0s
SensitivityCalibration 2 1 1 25s
WavelengthCalibration 2 1 1 39s

Session Figures

No session figures found for the loaded instrument/session.
No figures loaded

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
okay

SO2

Decision for SO2
Yes
no events visible but a nice flat background

HCHO

Decision for HCHO
With Care
noisy pattern around noon

H2O

Decision for H2O
Yes
very good aggremant with Aeronet