Decision Summary
Upload-ready calibration protocol overview
Instrument
P118s1
Session
86
Location
-
Calibration person
Hanna Rohringer
CF file
Pandora118s1_CF_v86d20170202.txt
Software version
1.8
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: With CareO3Synt: YesNO2: YesSO2: With CareHCHO: UnusableH2O: Yes
Action Timeline
15 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 1m 3s |
| CreateReference | 2 | 1 | 1 | 3s |
| FieldCalibration | 22 | 12 | 2 | 2h 39m |
| No action | 14 | 0 | 0 | 0s |
Session Figures
16 figures from Google Drive/figures
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora118s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 129 KBCreate reference SELSPECS
Pandora118s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 120 KBCreate reference SLERR (STRAY)
Pandora118s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 87 KBCreate reference AODEST
Pandora118s1_CreateReference_AODEST.jpeg - 46 KBCreate reference SENSCORRFIN (SENS)
Pandora118s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 104 KBCreate reference DISPFIT (DISP)
Pandora118s1_CreateReference_DISPFIT(DISP).jpeg - 85 KBCreate reference WLCHANGE (DISP)
Pandora118s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 134 KBCreate reference SHIFT (CLOS)
Pandora118s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 117 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
out was a mess with nearly every ref day - adapted cfref for double diff manually. , ExtRef with cate - high straylight level
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
Yes
SO2
Decision for SO2
With Care
left out uncalibrated as MLE was not applicable in this period
HCHO
Decision for HCHO
Unusable
Delrin parts
H2O
Decision for H2O
Yes