Decision Summary
Upload-ready calibration protocol overview
Instrument
P115s1
Session
5
Location
ISACROMA
Calibration person
Moritz Mueller
CF file
C:/Blick/data/calresults/Pandora115s1/session5/Pandora115s1_CF_session5.txt
Software version
1.8.58
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: YesO3Synt: With CareNO2: YesSO2: UnusableHCHO: With CareH2O: Not answered
Action Timeline
8 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| CreateReference | 152 | 5 | 2 | 11m 27s |
| HealthCheck | 2 | 1 | 1 | 2s |
| No action | 6 | 0 | 0 | 0s |
| StrayLight | 4 | 2 | 2 | 7s |
Session Figures
No session figures found for the loaded instrument/session.No figures loaded
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
ok no interpolation needed
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
ok, E-MLE used
SO2
Decision for SO2
Unusable
HCHO
Decision for HCHO
With Care
most days look good, some are unusuable, E-MLE used
H2O
Decision for H2O