Decision Summary

Upload-ready calibration protocol overview
Instrument P115s1
Session 4
Location ISACROMA
Calibration person Moritz Mueller
CF file Pandora115s1_CF_v4d20170120.txt
Software version 1.8.26
Calibration vector -
Reference day -
Overall status development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Not answered
Instrument P115s1
Session 4
Events 505
Figures 0
Action time 39m 40s

Action Timeline

28 action intervals
Nov 26, 2018Nov 29, 2018Dec 12, 2018Dec 13, 201809:51 AMCheckDispersionCreateReferenceAssignCalFileCheckDispersion 2m 15s finishedCheckDispersion 25s finishedCheckDispersion 0s finishedCreateReference 3m 19s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCheckDispersion 3s abortedCreateReference 4s abortedCreateReference 2m 25s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 6s abortedCreateReference 7s abortedCreateReference 24m 34s finishedCreateReference 0s finishedCreateReference 0s finishedAssignCalFile 52s finishedAssignCalFile 22s finishedCreateReference 19s abortedCreateReference 3m 7s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 8s abortedCreateReference 0s staleCreateReference 0s finishedCreateReference 0s finishedCreateReference 1m 34s aborted

Duration by Step

CreateReference35m 43sCheckDispersion2m 42sAssignCalFile1m 14s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 8 2 0 1m 14s
CheckDispersion 12 4 1 2m 42s
CreateReference 469 21 8 35m 43s
No action 16 0 0 0s

Session Figures

No session figures found for the loaded instrument/session.
No figures loaded

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
no extrapolation needed, unct=5e-3

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes

SO2

Decision for SO2
Yes

HCHO

Decision for HCHO
Yes

H2O

Decision for H2O